ENMET - Model MedAir 2200 -Compressed Airline Monitor
MedAir 2200 is a compact, easy to install and operate compressed airline monitor that can continuously and simultaneously monitor up to four gases in real-time using internally mounted sensors. MedAir 2200 is typically configured to monitor hospital compressed air for carbon monoxide, oxygen and dew point.
The MedAir 2200 is a versatile microprocessor-based compressed air line monitor for medical air systems. NFPA-99 requires continuous monitoring of carbon monoxide and dew point in medical compressed air lines. The MedAir 2200 configured to monitor CO and dew point is designed to meet these requirements.
The system can be custom configured to detect any combination of the following hazards: carbon monoxide, dew point, oxygen, and a number of other harmful gases including carbon dioxide. The instrument has a user-friendly interface for all maintenance and operation functions and is provided in a convenient package for hospital air line monitoring applications.
- Carbon Monoxide and Dew Point Monitoring in One Package
- Simple Push-Button Calibration Procedure
- Meets NFPA-99 Requirements
- Programmable Relay Contacts
- Large Easy-to-Read Display
- Alarm Points Adjustable within Preset Limits
- 4-20 mA Output
- RS-232 / RS-485 Modbus Communication Capability
- Many Instrument Configuration Options Available for CO, Dew Point, Oxygen, Carbon Dioxide and Other Toxic Gases
- Approved to UL 60601-1, IEC 60601-1
CSA 22.2, No. 601.1
- Operating Power : 100 to 240 VAC and/or 12 VDC, 15 Watts
- Display : 2 Line, 16 Character, Dot Matrix LCD
- Monitor Type : CO and Dew Point
- Visual Alarm : LEDs
- Audible Alarm : Piezo Electric
- Alarm Relay : 5 programmable relays plus fault. All relays are programmable latching or non-latching, dry DPDT, 10 amps (resistive load only) at 110 Vac.
- Temperature Range Recommended : 35 to 77 °F (0 to 25 °C)
- Sensors : CO: Electrochemical Dewpoint: Thin-Film Polymer
- Sensor Life : CO: 1-3 years Dewpoint: 2 years, replacement
- Response Time : CO: T90 = 30 seconds
- Dewpoint: T90 = 10 seconds for -40° to 50°F step change
- Inlet Pressure : 55 psi facility Air Supplied
- Enclosure : Thermoplastic box with clear, hinged front cover, designed for NEMA 12 and 4X.
- Size : 10.5 x 8.5 x 5.8 in
- Weight : 8 lb
- Wiring : Board Mounted Terminal Strip
- Horn : 95 dB at 2 feet
- Flow Rate : 1 liter per minute (2 SCFH)
- Standard Calibrations
- Alarm Point : Range
- 10 and 20 ppm CO : 0-50 ppm CO
- +35º F and +50º F Dew Point : -112º to +68º F Dew Point
- All alarm points are adjustable
- Optional Configurations
- MedAir 2200 P/N 03420-000 for monitoring CO only
- MedAir 2200 P/N 03420-001 for monitoring CO and Oxygen
- MedAir 2200 P/N 03420-006 for monitoring CO, Oxygen and Dew Point
- MedAir 2200 P/N 03420-008 for monitoring CO, Oxygen, Dew Point and CO2