DitecoModel 1812 -Laboratory Moisture Calibrator

SHARE

As calibration cylinders of known moisture content are unstable and cannot be prepared with any accuracy or repeatability, Diteco developed an instrument capable of producing a known concentration of water in a carrier gas, typically nitrogen. It can be used to calibrate other moisture monitors, or as a stand-alone moisture monitor.

Most popular related searches

The operating principle of the calibrator is to generate a continuous flow of low level (ppm) water vapour in a nitrogen carrier gas to be used to calibrate other moisture monitors. This device can also be used as a moisture analyser if required.

Dry nitrogen is fed through a bubbler, the flow rate being adjusted to deliver a constant flow of known ppm moisture in the stream. An internal analyser measures the moisture content and the instrument delivers up to 2 l/min moisture in nitrogen to an external moisture monitor which in turn can be calibrated or validated.

  • Suitable for calibrating or validating any moisture monitor in the 0-1,000 ppm by volume range.

ANALYSER:

  • Ranges: 0-1,000 / 0-10,000 ppm(v)
  • Accuracy: ±1% of range at 20°C
  • Ambient: 10°C to420°C
  • Recorder Output: 4-20mA
  • Alarm: One fully adjustable relay output
  • Calibration Sample Flow Output: up to 2 l/min
  • Dimensions: 365(w), 192(h), 360(d) mm
  • Weight: 7 Kg

UTILITIES

  • Power: 100-120, 220-240 VAC

CARRIER GAS

  • Pressure: 0.4 to 4 Bar g maximum
  • Type: Dry nitrogen
  • Flow Rate: up to 2.5 l.min
  • Options: Fault alarm
  • Low flow alarm