PID Analyzers, LLC
  1. Companies
  2. PID Analyzers, LLC
  3. Products
  4. PID Analyzers - Model 1000 - ...

PID AnalyzersModel 1000 -Electrochemical Analyzers for Toxic Gases in Air

SHARE

The Model 1000 is a low cost continuous Analyzer. There are a number of sensors 30+ EC, CG & TCD. It has an optional 4-20 mA output, a single programmable setpoint, larger pump & an industrial NEMA 4 enclosure. This 1000 has a variety of sensors to chose from (>40) and comes with an Operator`s Manual, configuration software and RS232 cable.

Most popular related searches
  • Single piece construction : The single piece construction NEMA 4 enclosure constructed of fiberglass is waterproof, very rugged and durable. An optional 19” rack mount enclosure is also available
  • Ease of calibration : A two point calibration is used for all detectors. A zero gas and a span gas are needed. The time to calibrate varies from about 20 seconds to 2 minutes. The Cal button is located on the main menu for ease of use.
  • Library : There are libraries for the CG and TCD sensors that are easily accessible.
  • Display : Our 2 line by 16 character display is backlit. The upper display provides the concentration and the bottom display provides the sensor, units, alarm and logging mode. For multisensor units, a 3 readout LED is used.
  • ADC : Our 16 bit ADC provides a resolution of 1 part in 64,000 and our signal algorithm minimizes noise by signal averaging in the msec range./td>
  • Programmable Voltage/current output : 0-1VDC/4-20 mA.
  • Detectors Available: Electrochemical (EC) & CG or TCD.
  • Sensor(s): Single or up to 3 EC sensors plus 1 TD or 1 CG
  • Span Drift: <2% over 24 hours
  • Power: 90-220 VAC or 24 VDC
  • Performance: Depends on sensor.
  • EC Sensors: CO2, CO, O2, O3, H2O2, SO2, H2S, THT, NO, NO2, NH3, AsH3, PH3, SH4, HCl, Cl2, ClO2,F2, HBr, HF, HCN, I2, Phosgene, H2, ETO, Formaldehyde, C2H4, 1,3 BD, Acrylonitrile
  • Precision: +/- 2-3% of reading
  • Range: Depends on sensor see brochure.
  • Dimensions: 6.75`` W x103/8``H x 6 3/4``D
  • CG Sensor: Hydrocarbons, H2, NH3…
  • Response time: <10 sec. to < 60s FID (to 90%) response
  • Linear Range: See brochure.
  • Weight: 7.4 pounds.
  • TCD Sensor: H2, He, Xe, CO2, …
  • Sample flow rate: > 0.2 LPM
  • Power consumption: < 1 watt maximum.
  • Instrument operating conditions: 5-40 degrees C, 0-95% RH (non-condensing).