ENVEA - Model MC 3000 -Generation of Mercury Calibration Gas
FromENVEA
The MC-3000 generates a continuous stream of a mercury vapor loaded gas in order to check or calibrate mercury analyzers. An exactly adjustable carrier gas stream (normally air or nitrogen) is loaded with mercury in a first step. In a second step the gas is supersaturated by cooling. Subsequently the excess mercury is condensed in a special cell to get exact saturation. The gas is then diluted, the dilution ratio is continuously adjustable in order to obtain concentrations within a specific range.
Most popular related searches
calibration gas
calibration gas flow
mass flow control
mercury calibration gas
mercury calibration
gas calibration
mercury vapor
mercury absorption
mass flow
ambient gas
- Generates elemental mercury vapor
- Continuous operation
- Approx. 10 to 500 µg/m³ Hg°
- Compact sized and portable
- No permeation principle – no need to calibrate the calibrator
- Microprocessor controlled
Suitable for all applications where a gas stream with a preset and constant mercury concentration is needed.
- Range of concentrations: approx. 10 to 500 µg / m³ Hg°
- Generated calibration gas flow: 1 to 9 l/min, depending on conc.
- Carrier gas: Air, Nitrogen
- Carrier gas flow: max. 12.5 l(min)
- Carrier gas pressure: 1 to 3 bar
- Cleansing of carrier gas: builtin particular filter (0.2 µm )
- Mercury absorption filter: sulphurized activated carbon at outlet
- Temperature sensors: Pt-100 with 0.01 °C accuracy
- Flowmeter: electronic mass flow controller (MFC)
- Connectors: for tubing 4mm i. d. / 6mm o. d.
- Power supply: 230 VAC / 50 Hz (110 VAC / 60 Hz option)
- Power consumption: 30 VA
- Dimensions (W x H x D): 45 x 18 x 27 cm
- Weight: approx. 5 kg
- Our products are manufactured according to the ISO 9001 quality regulations.
- CE approved