ENVEA - Model MIR 9000 CLD -Multi-Gas NDIR-GFC Analyzer
FromENVEA
Standard Reference CLD method for low & ultra low NOx measurement, IR-GFC for CO, CO2, SO2, N2O, Nox, HF, HCl, TOC and O2 in a single analyzer.
Most popular related searches
multi-gas
cold sampling system
multi-gas IR
gas filter correlation
multi-gas analyzer
EN 15267
gas analyzer
IR gas analyzer
gas cooler
unheated sampling
Multi-Gas IR-GFC analyzer (Infrared Gas Filter Correlation) with CLD option (Chemiluminescence Detector) designed to measure dry and corrosive samples
- Over 4000 installations worldwide, covering many applications and industries
- Unheated sample line thanks to the permeation drying technology used
- Intrinsic security with on-board residual H2O measurements
- On-board oxygen measure for environmental reporting
- Interactive menu-driven software with real time graphic display for ease of operation
- Highly accurate, excellent stability with automatic optical stability check
- Automatic CO2 and cross interference correction
- Built-in data logger for 7 additional parameters (flow, pressure, temperature or any other analogue input)
- Compatible with various drying technologies such as a SEC sampling system or a high performance gas cooler
- Cogeneration, Gas Turbines
- Industrial Boilers
- Furnaces
- Power & Combustion
Measurable parameters
Pollutant: Range
- HCI: 0-15 / 5000
- HF: 0-25 / 300
- NO: 0-20 / 2000
- NO2: 0-20 / 2000
- N2O: 0-20 / 2000
- NOx: 0-20 / 2000
- SO2: 0-75 / 5000
- CO: 0-75 / 10000
- CH4: 0-10 / 1000
- THC: 0-50 / 5000
- CO2(%): 0-10 / 100 or 0-20
- O2(%): 0-10 / 25
Lowest / Highest available ranges expressed in mg/m3 (may vary with your site conditions to be indicated on the Site Survey Form)
Technical Specifications
- QAL1 to EN 15267-3 certified: YES
- US EPA compliance: YES
- Sampling technology: Cold/Dry Extraction
- Sample transfert: non heated polytube
- back flush function: YES
- span injection: YES on analyser or at sample point
- Built-In O2 option: YES
- Serial link communication: RS232 / RS422
- Other Communication: Ethernet
- Analogue OUTPUTS: Optional
- Analogue INPUTS: YES (x 4)
- Dimensions (mm): • Tight box: 200 x 600 x 600 (D x W x H)
- • 19” rack: 490 x 483 x 177 (D x W x H)
Complete system would normally comprise of:
- Sample extraction and conditioning probe (with integrated temperature, pressure and flow measurement)
- Cold sample lines
- Calibration module
- Instrument air drying module
- Acquisition and data management system (DAHS) may complement the system
- COMPLIANCE WITH LCPD AND WID APPLICATIONS
- QAL 1 certified by the MCERTs to EN 15267-3
- QAL3 compliance as defined by EN 14181
- U.S. EPA approved as compliant with 40 CFR 60 ET 75