The mercury analyzer PA-2 is used for continuous monitoring of mercury concentrations in industrial processes. The PA-2 was specially designed for process analysis and operation under harsh industrial condition. The system comes with either the VM-3000 or the UT-3000 photometer to meet different requirements regarding accuracy and sensitivity of measurements.

Most popular related searches
  • Fully automatic system
  • Easy menu-driven operation
  • Proven and reliable detection method: AAS
  • Measuring ranges from 10 ppt to 100 ppm
  • High flexibility of sample pretreatment
  • Suited for complex sample compositions
  • Low reagent consumption
  • Corrosion-protected construction
  • Automatic self diagnosis system for reliable operation

  • Quality control in chlorine-alkali plants
  • Monitoring of scrubber water of waste incinerators and power plants
  • Control of industrial sewage
  • Control of purification plants
  • Quality control of sulphuric acid
  • Quality control of drinking water
  • Measuring principle; Amalgamation on MI GoldTrap (PA-2 Gold only), cold vapor atomic absorption (CVAAS) at 253.7 nm
  • UV-Source; Electrodeless low pressure mercury lamp (EDL)
  • Stabilization method; Reference beam method
  • Optical cell; Fused silica (Suprasil), lapprox. 230mm, heated approx. 55°
  • Measuring ranges PA-2; 0.1 … 1 µg/l to 0…10 mg/l
  • Measuring ranges PA-2 Gold; 0.01 … 1 μg/l to 0.1 … 10 μg/l
  • Sensitivity; > 0.01 µg/l
  • Response time; approx. 1 minute
  • Sample flow; approx. 20 l/h
  • Carrier gas; compressed air, filtered, approx. 30l/h, 1bar
  • Reduction reagent; Tin-II-chloride or Sodiumhydroborate
  • Reagent consumption; approx. 3l/week for each reagent
  • Sample digestion; HCl or H2SO4, KMnO4 or H2O2 or Fenton’s Reagent or NaClO3; depending on sample composition
  • Signal outputs: analogue: 4 – 20 mA (500 Ω max.);
  • serial: RS232 / USB/td>
  • Status output: Calibration; Autozero; Service; Malfunction; according to NAMUR 64 recommendations
  • Housing: Fibreglass-reinforced polyester, for use in highly corrosive environment – Reagent storage cabinet made of chemically resistent plastic material/td>
  • Protection class: IP 66 (EN 60529/ NEMA 3; 3R; 4; 4X; 12; 13)
  • Operating temperature: 0 to 40°C
  • Sample temperature: 65°C
  • Power supply: 230 VAC / 50 Hz (110 VAC / 60 Hz option)
  • El. Power consumption: 750 VA max. (PA-2 Gold, heating peak)
  • Dimensions (W x H x D): 62 x 78 x 33 cm
  • Weight: approx. 50 kg
  • Our products are manufactured according to the ISO 9001 quality regulations.
  • CE approved