Envirotech Instruments Pvt. Ltd. (EIPL)

EnvirotechModel APM 460DXNL -Advanced PM10 Samplers Used for Precise Monitoring

SHARE

APM 460DXNL is one of the most advanced PM10 samplers manufactured in India. It is based on CSIR NEERI technology and is the only PM 10 sampler in India that conforms to BIS 5182 (Part 23): 2006.

Most popular related searches
  • Automatic Flow Controller with electronic feedback for constant sampling rate throughout the sampling period.
  • Brushless regenerative blower with very little noise.
  • Heavy duty Induction motor fitted in blower requires no maintenance.
  • The blower is immune to voltage fluctuations thus eliminating the need for a voltage stabilizer.
  • Electromagnetic interference (EMI) to TVs is eliminated.

Automatic flow controller and auto shut off

In all previous models of particulate samplers in use in the country, the sampling rate varies with dust loading on the filter and fluctuates with changes in the mains voltage. The APM 460DXNL uses an electronic flow controller to automatically maintain a constant sampling rate through the filter/cyclone assembly. The size fractionating cyclone of the APM 460DXNL thus operates close to its designed separation velocity resulting in a sharper cut-off for particulates larger than 10 microns. When the dust load on the filter becomes excessive and exceeds the capacity of the blower the electronic circuits provided in the flow controller automatically shut down the machine.

Owing to its modular design, APM 460DXNL can be easily paired with a gaseous sampling attachment (for monitoring SO2, NOx, NH3, Ozone, etc) as gaseous sampling requires only a few LPMs of airflow. This is possible through an attachable subsidiary unit APM 411 or the more modern APM 411TE. Kindly refer to the respective brochures of APM 411 and APM 411TE for details

  • Flow Rate: 0.8 – 1.4 m3/min free flow with flow stabilization by Electronic Feedback Flow Control Device.
  • Flow Rate Control: The system will automatically maintain the flow rate constant through an Electronic Flow Controller with pressure sensor and compensate for dust load deposited on the filter. When the dust load increases beyond the capacity of the blower the flow rate will begin to decrease and the system will automatically be shut-down if the flow rate tends to drop below 0.85 m3 /min (As recommended in BIS IS 5182-Part 23:2006).
  • Particle Size: A cyclone is used for fractioning the dust into two fractions at D-50 cut off for 10 microns. Particles from 10 micron down to 0.1 micron are collected on Filter Paper placed on filter holder. Coarse particles (bigger than 10 microns) are collected in a cup under the cyclone.
  • Sampling Time: 28 hours (maximum)
  • Sampling Time Record: 0 to 9999.99 hrs. recorded on a Time Totalizer
  • Power requirement: operated at 220V AC (+/-) 10V single phase 50 Hz, supply continuously for at least 28 hours. For proper operation and safety a good earth connection is mandatory
  • Size & Weight: 400 x 300 x 650mm., 60Kg
  • Automatic Sampling: 24 hrs programmable timer to automatically shut off the system after pre-Control set time interval
  • Warranty: One year