- Home
- Companies
- F&J Specialty Products, Inc
- Products
- F&J - Model Displayless - PC ...
F&J - Model Displayless -PC Interfaceable Air Flow Calibrator
Notable Features: PC Interface program: Western or Asian Languages. Operator Selectable Options: Multiple language options; European or Asian Version. Multi gas option. Volumetric or mass flow. At Reading or Full Scale Accuracy Calibration Type. Reference T and Reference P. Engineering units for all reported parameters. Semi-Automatic calibration-verification when used with F&J digital air samplers. Customizable calibration report. Accuracy: ± 1.0% Full Scale (F.S.). Two Year Warranty. Complementary Storage Case.
Most popular related searches
air flow calibration
flow calibration
digital air sampling
differential pressure sensor
barometric pressure
mass flow
multi-gas
pressure sensor
flow sensor
air sampling
MEASUREMENT PRINCIPLE
- Flow Sensor: Differential Pressure Sensor
- Standardization: Operator selectable values for reference temperature and pressure
- Flow Curve linearization: Individually calibrated and software corrected
MEASUREMENT RANGES
- Air flow: Various flow ranges available
- Temperature: -40º - 122º F (-40º - 50º C)
- Barometric pressure: 30 - 22 In-Hg (760 - 559 mm-Hg); (101.325-74.5 kPa)
- approx. Sea level to 5900 ft. (1800 m) elevation above sea level
- Optional low range to 10 In-Hg (254 mg) (33.86 kPa)
MEASUREMENT ACCURACY
- Air flow: ± 1.0% of full scale
- Temperature (Typical): + 1ºC over range –20ºC to 50ºC
- Barometric pressure: ± 1% of reading over measurement range of 22.00-30.00 “Hg
STANDARD TEMPERATURE and PRESSURE CHOICES:
- Temperature 32ºF (0ºC) 59ºF (15.0ºC) 68ºF (20.0ºC) 70ºF (21.1ºC) 77ºF (25.0ºC)
- Pressure
- 29.92”Hg (1 atm, 760mm Hg, 1.013 bar, 101.325 kPa)
- 1 bar, (100kPa, 750mm Hg, 0.987 atm 29.53”Hg)
