PID Analyzers - Model FI53 -Flame Ionization Detector (FID)
The flame ionization detector (FID) is used for the selective measurement of any organic species that can be ionized in a flame. This corresponds to most hydrocarbons and is described below: The Model 53 has an FID (burner with HV bias and collector), a power supply for the FID bias, an electrometer and gauges and flow control for the support gases for the FID. The electrometer output is 0-10 VDC. The customer needs a separate ADC channel to feed the signal into the GC’s data system. Alternately, PID Analyzers sells a data system (Model 53) that will feed the signal into a separate PC. The FID is a sensitive and very selective detector for hydrocarbons. It has the wide dynamic range (>106).
Most popular related searches
flame ionization detector
flame ionization
flame detection system
hydrocarbons monitoring
laboratory instrument
flow control
FID detection
chromatograph
chromatography separation
chromatography
- Easy to install
- Adaptability: Is readily adaptable to any chromatograph.
- Selectivity: Improves separations and analysis of hydrocarbons.
- Used by environmental agencies worldwide
- Wide linear dynamic range: > 5 x 106
- Display: Our 2 line by 16 character display can be backlit; used for range and background readings.
- High sensitivity: pg detection limits.
- ADC: Our 16 bit ADC provides a resolution of 1 part in 64,000 and our signal algorithm minimizes noise by signal averaging in the msec range.
- Programmable Voltage output: 0-10VDC
- Species measured: Selective for Hydrocarbons
- Detector: > --`` D x ---`` L
- Detection limit: <20 ppb of benzene
- Electronics enclosure (FI53)
- Weight: 3 lbs.
- Size: - 7`w x 8.5``D 3.75`` H
- Range: > 2 x 106
- Power consumption: < 1 watt maximum
- Response time: <1 sec. to 90% response
- Instrument operating conditions: 5-40 degrees C, 0-95% RH (non-condensing)