SabioModel 2010D -Gas Dilution Calibrator (Portable, Rack or Bench)

SHARE

The 2010D Intelligent Gas Dilution Calibrator can easily perform gas dilution, multi-source gas blending, ozone and gas phase titration (GPT) for environmental, laboratory, or point source monitoring.

Most popular related searches

The Model 2010D represents the latest generation of calibrator technology for environmental, laboratory or point source monitoring. The unit has two internal mass flow controllers (MFC`s), with options for an internal Ozone generator and a photometer (5.25 inches high and weighing less than 25 lbs).

The Model 2010D calibrator uses embedded microprocessor technology to precisely deliver and control gas concentrations along multiple curve fits to linearize the desired outputs for the intended purpose. The unit can be operated manually, automatically, or semi-automatlcally by remote access. It also has the capability to produce and store in memory 20 calibration sequences with up to 20 levels of source/dilution in each sequence.

Users can choose either the front panel membrane keypad, external keyboard, serial port, or Ethernet to input calibration sequences.

The instrument`s pneumatic system uses Mass Flow Controllers to precisely meter gas streams and implements multiple types of curve correction algorithms to linearize calibrations. A second source MFC can be added for a wider dynamic range or trace level applications. An optional second source MFC with blending allows user-performed interferent testing and evaluation.

Calibrator design exceeds US EPA calibration method requirements. Dilution components are calibrated with standards and test equipment traceable to the National Institute of Standards and Technology.

Pnuematics

  • Diluent Mass Flow Controter. 0-10 SLM
  • Source Mass Flow Controller. 0-100 SCCM
  • 4 Calibration Gas Input Ports and 1 Purge Port
  • 5 Output Ports and 1 Vent Port
  • 1 Diluent Gas Input
  • Bright Active Matrix Color Display

Electronics

  • 42-Button Membrane Keypad
  • 8 Bits Digital Input`Output
  • 1 S«rial Port for Data Communications
  • External PC keyboard Input Port
  • Centronics Paratel Printer Port
  • Air Source Control Port

Electrical

  • Standard 93-264 VAC. 150-300 VA, 50/60 Hz
  • Cp-era`.lon
  • Alternate Diluent Mass Flow Controller Ranges 0-20 SLM
  • Alternate Source Mass Flow Controller, Ranges 0-2 SLM
  • Second Source Mass Flow Controller, Ranges 0-2 SLM
  • Additional Calibration Gas Input Ports
  • Permeation Port
  • Additional Output Ports
  • Second Diluent Gas Input Port
  • Internal Ozone Generator w/Gas Phase Titration (GPT) compatibility
  • Ozone Generator UV Optical Servo Control Loop
  • Internal Ozone UV Absorption Photometer
  • Multi-Source Gas Blending
  • Master Output Solenoid
  • Instrument Solenoid Driver for External Valves
  • Rack Mount Kit
  • Glass chamber for GPT
  • Ethernet TCP/IP
  • Transport case

Dilution System:

  • Input Dilution Gases: 1 Standard, 1 Additional (Optional)
  • Input Source Gases: 4 Standard, 1 Purge, 2 Additional (Optional)
  • Dilution Mass Flow Controller: 0-10 SLPM, Optional Range 0-20 SLPM
  • Source Mass Flow Controller: 0-100 SCCM, Optional Range 0-2 SLPM
  • 2nd Source Mass Flow Controller: 0-2 SLPM
  • Flow Accuracy: less than ±1% full scale
  • Flow Repeatability: less than  ±0.15% full scale
  • Linearity: less than  ±0.5% full scale
  • Response Time at Output: less than  1 minute

Optional Internal Ozone Generator:

  • 2ppb - 1000 ppb, Max - 6 ppm Flexcal
  • Accuracy: ± 1% of Set Point or ± 2 ppb @ 5 SLPM
  • Nominal Flow: 100 SCCM, ±1 SCCM
  • UV Lamp Temperature: 50°C,±0.1°C

Optional UV Absorption Photometer:

  • 0.1 ppm to 0 - 10 ppm (selectable)
  • Linearity: 1 Liter (nominal)
  • Precision: ± 1 ppb
  • Zero Drift: less than ± 1 ppb for 24 hours or 30 days
  • Flow Rate: 1 Liter (nominal)
  • Analog Inputs: 3 Ranges:   100 mV, 1V, 5V

Calibrator Interface:

  • Operation: Keypad, keyboard, serial, USB, Ethernet
  • Calibration Definitions: 20 factory-defined calibration sequences (less than 20 pts each)
  • Calibration Types: Gas Dilution, Ozone, GPT (Gas Phase Titration)
  • Gas Definitions: 2 Diluent Gases, 20 Source Gas
  • Auto Calibrations: 20 timer driven routines perform user-defined calibration sequences on a daily schedule
  • Digital Inputs: 24 status I/O bits for calibrator functions
  • Digital Input Types: Contact closure or TTL logic
  • Digital Outputs: 24 status I/O bits
  • Communications: Ethernet TCP/IP, RS-232, RS-485, USB

Physical & Operational Specifications

  • Operating Temperature: 5°Cto40°C
  • Dimensions: 5.25" (13.3 cm) H x 16.2" (41.2 cm) Wx 19.25" (48.9 cm) D
  • Average Weight: 24 lbs. (10.9 kg)
  • Input Voltage: 98-264 VAC, 150-300 VA, 50/60 Hz