- Home
- Companies
- Hiden Analytical Ltd.
- Products
- Hiden - Model CO-A - Carbon Monoxide ...
Hiden - Model CO-A -Carbon Monoxide Analyser
A CO Analyser for integrating with Hiden HPR-20 Gas Analysis systems complimenting mass spectrometer data for quantitative analysis of complex gas mixtures. Designed for gas analysis applications of syngas, biogas and catalysis studies where levels of CO are analysed by the CO-A in the presence of nitrogen, carbon dioxide, methane and other gas mixtures.
A CO Analyser for integrating with Hiden HPR-20 Gas Analysis systems complimenting mass spectrometer data for quantitative analysis of complex gas mixtures.
Designed for gas analysis applications of syngas, biogas and catalysis studies where levels of CO are analysed by the CO-A in the presence of nitrogen, carbon dioxide, methane and other gas mixtures.
Gas sampling and data collection from the CO-A are fully integrated with the mass spectrometer system in hardware and software.
A non-condensing gas sampling flow rate of ~ 800 ml/min is recommended for a response time of ~ 60 seconds.
The CO-A is available with two concentration measurement range options: 0 – 2%, and 0 – 10%
*Standalone CO Analysers with measurement ranges: 0.04 – 1,000 ppm and 0 – 10,000 ppm are also available, with data acquisition integrated to the Hiden mass spectrometers.
- Detection range options: 0-2% and 0-10%
- Limit of detection = 2% of full scale
- Response time: 60 secs to 95% response at 800 ml/min – non-condensing
- Accuracy: +/- 2% FS
- Repeatability: < 0.3% Rd
- Reproducibility: < 1.1% Rd over 5 days
- Rack mountable
- Standalone CO Analysers with measurement ranges: 0.04-1,000 ppm and 0-10,000 ppm available, with data acquisition integrated to the Hiden mass spectrometers.