Hiden Analytical Ltd.
  1. Companies
  2. Hiden Analytical Ltd.
  3. Products
  4. Hiden - Model HPR-60 MBMS - Molecular ...

HidenModel HPR-60 MBMS -Molecular Beam Mass Spectrometer System for Analysis of Neutrals, Radicals and Ions

SHARE

The Hiden HPR-60 molecular beam mass spectrometer is a compact skimmer inlet MS for the analysis of atmospheric plasma and reactive gas phase intermediates.

Most popular related searches
Radicals are sampled via a multistage differentially pumped skimmer inlet and transferred to the MS ion source with minimal interaction with other species and without wall collisions. Customisable inlets allow connection to many different reactor systems, including atmospheric plasmas.

The skimmer system, combined with a Hiden triple filter precision mass spectrometer, offers a sampling system with ultra fast response and high accuracy.

  • Molecular beam sampling at atmospheric pressure
  • +ve and -ve ion analysis
  • User replaceable skimmer cones (can be biased)
  • Electron attachment ionisation mode for the study of electro-negative radicals
  • APSI-MS soft ionisation mode for radicals analysis
  • Mass range options: 300, 510 or 1000 amu
  • Energy range options: 100 eV or 1000 eV
  • Mass Range: 50, 300, 500, 1000 or 2500 amu
  • Sampling Pressure Range: 10–4 mbar up to atmosphere depending on chosen configuration
  • Operating modes:
    • Threshold Ionisation
    • Electron Attachment Ionisation
    • Positive Ion Detection
    • Negative Ion Detection
    • Multiple Ion Detection
    • Bar Scan, Profile Scan
    • Ion Energy Distributions
  • Mass Spectrometer: Precision Triple Filter Quadrupole
  • Detector:
    • Off axis Positive and Negative Ion Pulse Counting Single Channel
    • Electron multiplier, 7 decade continuous dynamic range
  • Ion Source: Low profile electron impact with dual filaments
  • Energy Filter: Pole Bias, Bessel Box or 45 degree sector field options
  • Inlet:
    • Modular 2 or 3 stage close coupled skimmer cone orifice inlets with fine alignment of mass spectrometer across the molecular beam axis
    • Complete turbomolecular pump sets and vacuum gauging to suit each inlet
  • Data Acquisition and Control:
    • 32 bit MASsoft operating system with HAL 5 Microprocessor
    • Rotating disc chopper assembly for automated beam foregroundbackground measurement with beam chopping rates to 360 Hz.
    • Programmable Signal Gating Module providing full software control of beam chopper and timing circuitry in pulsed applications
    • 2 Analog Inputs, 3 Relay Outputs, 5 Digital I/O lines