- Home
- Companies
- Fluke Calibration
- Products
- Fluke - Model 5080A - High Compliance ...
Fluke - Model 5080A -High Compliance Multi-Product Calibrator
The 5080A Multi-Product Calibrator calibrates your analog and digital workload accurately and economically. Its high voltage and current compliance makes analog workload calibration easy and precise. And built-in protection circuitry protects it against damaging input voltages.
Calibration solutions for your analog and digital workload
- High compliance
- Protection circuitry
- Calibrates a wide workload, including analog meters and 3.5 and 4.5 digits DMMs
This easy-to-use multifunction calibrator can handle a wide meter calibration workload that includes:
- Analog meters
- Panel meters
- Digital multimeters
- Watt meters
- Clamp meters (with coil accessory)
- Megohm meters (optional)
- Oscilloscopes to 200 MHz (optional)
- …and more
Versatile software applications enable you to record paperless results, and more. Best of all, the 5080A Multi-Product Calibrator offers this versatile functionality at a price that fits nicely into your budget.
Accurate, reliable analog meter calibration
The 5080A Multi-Product Calibrator calibrates your analog workload accurately and reliably, thanks to its high voltage and current compliance. With maximum burden up to 800 mA for ac/dc voltage, and voltage up to 50 V for ac/dc current, 5080A calibrators can drive a wide range of analog meters.
Options and accessories expand workload coverage
Options and accessories enable you to use the 5080A Multi-Product Calibrator to calibrate an even broader workload, including:
- Clamp meters. The 9100-200 10/50 turn coil and 5500A/COIL 50-turn current coil enables the 5080A to calibrate most popular clamp meters at currents up to 1000 A rms amps.
- Oscilloscopes. Oscilloscope calibration to 200 MHz quickly, easily, and cost effectively. Verify dynamic response, bandwidth, timing, multiple triggering functions, and input resistance.
- Megohm meters. This option sources high ohms, high voltage resistors up to 18 G ohms. It also measures high voltage outputs.
Protective circuits prevent damage
Mains voltage inadvertently applied to a calibrator’s output terminals can cause extensive damage, requiring costly repairs. Electrical protection for calibrator outputs is vital for daily operation. The 5080A calibrator’s innovative protection circuitry prevents it from being damaged by reversed input voltage, so you can use it with confidence day after day.
Highest voltage and current compliance
The 5080A has the highest voltage and current compliance of any calibrator in the Fluke Calibration multi-product and multifunction calibrator families, making it an ideal solution for calibrating analog meters and other instruments requiring higher drive capability for proper operation.
- Voltage DC :
- Range: 0 to ±1020V
- Best 1 Year Specification: ±0.01% of setting
- Current DC :
- Range: 0 to ±20.5A
- Best 1 Year Specification: ±0.05% of setting
- Resistance :
- Range: 0 to 190 MO, 19 values in x1 and x1.9
- Best 1 Year Specification: ±0.025% of setting
- Voltage AC :
- Range: 1 mV to 1020V
- 45Hz to 1 kHz, Sine
- Best 1 Year Specification: ±0.1% of setting
- Current AC :
- Range:29 mA to 20.5A
- 45 Hz to 1 kHz, Sine
- Best 1 Year Specification: ±0.1% of setting
- DC Power :
- Range:10.9 mW to 20.5 kW
- Best 1 Year Specification: ±0.11% of setting
- AC Power :
- Range: 10.9 mW to 20.5 kW
- 45-65 Hz, PF=1
- Best 1 Year Specification: ±0.36% of setting
- Phase :
- Range: 0 to ± 179.99°
- 45 Hz to 1 kHz
- Best 1 Year Specification: ±0.25° of setting
- Frequency :
- Range: 45 Hz to 1 kHz
- Best 1 Year Specification: ±50 ppm of setting