Chromatotec
  1. Companies
  2. Chromatotec
  3. Products
  4. airmoPURE - Model D - High Purity Air ...

airmoPUREModel D -High Purity Air Generator

SHARE

We propose a generator of high purity air for our analyzers as well as for thedilution of permeation tubes. This air can be used as carrier gas for our`MEDOR` instruments as well as for FID detector.

Most popular related searches

The airmoPure provides zero air for :

  • FID detector: 180 ml/min
  • Pneumatic valve : 60 ml by commutation
  • Permapure : 150 ml/min
  • Flow in the cooling box ( Pelletier effect): 13 ml/min Dilution of permeation tube for calibration: 50mL/min in continuous mode
  • Carrier gas for MEDOR

The airmoPure includes :

  • A compressor fixed and a 4-liter cylinder at 7 bars which delivers outlet pressure at 3 bars.
  • A membrane air dryer with automatic purge .The air is dried before being compressed in the cylinder.
  • A heated catalyser to eliminate hydrocarbons at PPT level
  • Connections between the airmoPURE and the instrument
  • 180 ml/min for FID
  • 50 ml/min in continious for Calibration
  • Dilution: 230 ml/min in discontinuous mode (during calibration only)
  • Benzene permeation tube (32 ng/min ± 10% at 45°C) diluted at 230 ml/min = 139 μg/m3 or 37 ppb
  • Zero air for analysis in continuous mode (airmoBTX /CALIB): 180 + 50 ≈ 230 ml/min
  • Zero air for analysis in discontinuous (airmoBTX/CALIB): 180 + 230 ≈ 410 ml/min
  • Zero air total hydrocarbon concentration:
    • < 0.1 ppb NMTHC in ambient air without poltion
  • Dew point:
    • < -15°C°
  • Zero air maximum flow:
    • 1 l/min in discontinuous mode for airmOzone
  • Noise:
    • 55 dB(A)
  • Zero air start up time:
    • 45 min
  • Power supply:
    • 230V / 110V / 115V and 50 Hz/60Hz)
  • Electrical consumption:
    • 310 VA
  • Dimensions and weight:
    • Height: 370 mm
    • Width: 370 mm
    • Depth: 370 mm
    • Net Weight: 17 kg