The “Conductivity-Temperature Transfer Standard CT-01Plus” is a compact and rugged instrument designed for the metrology laboratories to perform calibration of oceanographic CTD probes. It combines a Primary Standard 100 ohm RPRT (Reference Platinum Resistance Thermometer), a large-sized Idronaut self-flushing cell provided with seven internal rings of fused platinum in the quartz with a microprocessor controlled high resolution and high accuracy electronic amplifier.

Most popular related searches

Optionally, the CT-01Plus can become a redundant measuring system when equipped with two temperature and two conductivity sensors.

The high performance electronics collects and transmits the temperature and conductivity readings through the RS232C interface from 1 to 10 times per second. The functions of the CT-01Plus are managed by means of a simple communication protocol through the RS232C interface. A Windows interfacing and calibration management programme “ICAL”, which greatly simplifies the CTD probes calibrations, completes the CT-01Plus.

  • Large-sized self-flushing cell provided with seven rings of fused platinum in the quartz.
  • High-performance 24-bit digitizing circuits.
  • Temperature controlled oven conditions the electronics to obtain drift-free measurements.
  • 1ppm Vishay resistors used as a reference for the CT-01Plus electronic preamplifiers.
  • Redundant C/T sensor pair.
  • Windows management software simplifies system operations and automates CTD calibration.

Components:

  • CT-01Plus Deck Unit.
  • RPRT Reference Platinum Resistance Thermometer.
  • Conductivity Cell.
  • Windows “ICAL” Management Software.

CT-01Plus Temperature

  • Sensor: Standard 100 ohm RPRT (Reference Platinum Resistance Thermometer).
  • Range: -5 to +45°C.
  • Initial Accuracy: 0.001 °C.
  • Resolution: 0.0001 °C.
  • Digital resolution: 0.00000596°C (23 bits).
  • Stability: +/- 0.0003°C/month.
  • Sampling rate: from 1 to 10 Hz.
  • Noise: <0.0003°C (1Hz).
  • Excitation current: 1.2 mA at 10 KHz.

CT-01Plus Conductivity

  • Sensor: Self-flushing cell provided with seven internal rings of fused platinum in the quartz (16 mm, diameter 110 mm length).
  • Range: 0 to 70 mS/cm.
  • Initial Accuracy: 0.001 mS/cm.
  • Resolution: 0.0001 mS/cm.
  • Digital resolution: 0.000008345 mS/cm (23 bits).
  • Stability: +/- 0.001 mS/cm/month.
  • Sampling rate: from 1 to 10 Hz.
  • Noise: < 0.0005 mS/cm (1Hz).
  • AC Excitation freq.: 10 KHz.

CT-01Plus Characteristics

  • Operating environment: Temperature: 10-25 °C.
  • Humidity: 10 to 90% no condensing.
  • Warm-up: 60 minutes.
  • Oven temperature: Typical operating temperature: 35°C.
  • Stability: +/- 0.1 °C.
  • Interfaces: RS-232C.
  • Communications: Byte oriented proprietary communication protocol and a menu driven interface for maintenance access.
  • Power supply: 115 or 230 VAC +/- 10%, 50/60 Hz.
  • Dimensions: 480 x 350 x 190 mm.
  • Weight: 10 Kg.

Calibration and management program that greatly simplifies and automates the calibration of conductivity, temperature and pressure sensors.

Interface up to 4 Idronaut OCEAN SEVEN 3xx CTD and the CT-01STD.