H2scan’s HY-OPTIMA 5000 Series analyzers utilizes H2scan’s Gen 5 technology to provide the most accurate, tolerant, and affordable hydrogen process gas measurement solution for industrial markets. The general purpose-rated analyzer uses a solid-state, non-consumable sensor for direct hydrogen measurement in process gas streams, with no cross sensitivity to other gases. Auto-calibration ensures the measurement stays within spec, even in dynamic conditions. Communication with the unit is via serial communication using Modbus RTU over RS485.
General Use Inline Hydrogen Process Analyzer
- Solid-state, non-consumable sensor
- No cross-sensitivity to other gases
- Automated calibration maintains measurement accuracy
The HY-OPTIMA models 5031, 5033, and 5034 analyzers are intended for use in dry gas streams where hydrogen is always present. The sensor can be safely exposed to hydrogen continuously. The model 5032 is for use in processes where hydrogen is occasionally or intermittently present, as may occur if there is a leak or an upset condition. The analyzers are CE approved for general use operation. H2scan’s HY-OPTIMA 5000 Series inline hydrogen process analyzer featuring auto-calibration is ideal for standalone use or OEM integration into existing analyzers. The sensors are used in gas streams where real-time, hydrogen-specific measurements can enhance process plant efficiencies, improve yields, reduce maintenance costs, and enable the hydrogen economy.
The HY-OPTIMA 5000 Series inline hydrogen process analyzer is ideal for standalone use or OEM integration into existing analyzers, in gas streams where real-time, hydrogen-specific measurements can enhance process plant efficiencies, improve yields, reduce maintenance costs, and enable the hydrogen economy.
Refining
- Catalytic reforming
- Hydrodesulphurization
- Tail gas treating units
- Flare monitoring
- Fuel gas
Natural Gas
- Wobbe Index or calorific value
- Blending and injection points
- Compressor stations
Hydrogen Economy
- Fuel cells and electrolyzers
Petrochemical
- Polymer feeds and flare gas process streams
Industrial Gas Supply and Hydrogen Production
- Air separation
- Steam methane reforming
HY-OPTIMA 5000 Series Specifications
Performance
- Operating pressure at analyzer:
- Recommended: 0.95 – 1.1 atm absolute (14.0 – 16.1 psia)
- Maximum: 7 atm absolute
- Note: Analyzers are factory calibrated at 1 atm abs.
- Process gas temperature: -40 to 60°C
- Flow rate: 0.1 to 10 SLPM (3/4″ tube)
- Operating humidity: < 95% RH (non-condensing)
- Calibration interval: None (auto calibrating)
Output Signal
- Digital: RS485, 2-wire; 19200 baud, 8-bit data, 2 stop bits,
- No parity; Modbus RTU protocol
- Analog: 4-20 mA available via optional module
- Relays: Four Form C relays available via optional module
Power
- Input voltage: 9 – 48 VDC
- Input power: 10 W
Physical
- Dimensions: 5.9 in (L) x 1.6 in (W) x 1.6 in (D)
- 15.1 cm (L) x 4 cm (W) x 4 cm (D)
- Weight: 0.8 lbs (0.4 kg)
- Electrical fitting: M12 4-wire
- Sensor fitting: ¾”-14 MNPT
Environmental
- Ingress protection: IP68
- Operating temp: -40 to 70°C
- Storage temp: -40 to 105°C