SPELEC RAMAN is an instrument for performing spectroelectrochemical RAMAN measurements. It combines in only one box a LASER Class 3B (785 nm), a Bipotentiostat/Galvanostat and a spectrometer (wavelength range 785 – 1010 nm and Raman shift 0-2850 cm-1) and includes a dedicated spectroelectrochemical software that allows optical and electrochemical experiments synchronization.

Most popular related searches
  • Current accuracy : ≤0.5 % of current range at 100 nA to 10 mA
  • Current resolution : 011 % of current output range
  • Dimensions in mm (W/H/D) : 25 X 24 X 11 cm
  • Light Source (LS) wavelength range : 785 ± 1 nm
  • LS Fiber optic connector : FC
  • LS Stability : ± 0.1 nm (-20 to 55 °C)
  • Maximum current in Ampère : ±40 mA
  • Maximum number of channels : 1
  • Maximum number of WE sharing AUX and REF : 2
  • Measured current resolution : 0.025 % of current range (1 pA on lowest current range)
  • Measured potential resolution : 0.012 % of potential range
  • Multichannel instrument : NO
  • Number of current ranges : 8
  • Number of current ranges remarks : ±1 nA to ±10 mA
  • Number of potential ranges (galvanostat) : 2
  • Number of potential ranges remarks (galvanostat) : ±100 mV to ±1 V
  • Operating modes : Bipotentiostat; Potentiostat; Galvanostat
  • Optical power output : 500 mW (375 mW typical)
  • Output power stability : ± 1 %
  • PC interface : USB
  • Potential accuracy : ±0.2 %
  • Potential range in volt : ±4 V
  • Potential resolution : 1 mV
  • Power : 5 V DC
  • Raman shift : 0 - 2850 cm-1
  • Spectral line width : <0.2 nm FWHM
  • Spectrometer (SPM) Detector : 2D CDD Array, Back thinned TE Cooled
  • SPM A/D resolution : 18 bit
  • SPM Dynamic range : 85000 : 1
  • SPM Fiber optic connector : SMA 905
  • SPM Integration time : 8 ms to 60 min
  • SPM Optical resolution : 7 – 11 cm-1
  • SPM Pixels : 1044 x 64
  • SPM Signal-to-noise ratio : 1000 : 1 (at full signal)
  • SPM Stray light : < 0.08 % at 600 nm; 0.4 % at 435 nm
  • SPM Wavelength range : 785 - 1010 nm
  • Weight of the instrument in kg : 3.4