John Caunt Scientific Ltd.
- Home
- Companies
- John Caunt Scientific Ltd.
- Products
- JCS - Model Alpha/Beta SiPM - ...
JCS - Model Alpha/Beta SiPM -Scintillation Detector Probe
Thin Alpha/Beta SiPM Scintillation detector: JCS is pleased to announce the new Scionix thin Alpha/Beta SiPM Scintillation detector with a built-in temperature compensated bias generator and preamplifier.
Most popular related searches
scintillation detector
alpha beta scintillation detector
beta scintillation detector
alpha scintillation detector
contamination monitoring
alpha contamination monitoring
- Pulse rise time (0-100%): 150 ns (betas) & 1μs (alphas)
- Pulse fall time (1/e): 1.2 μs (betas) & 3 μs (alphas)
- Gain: Approx. 75 mV/MeV(betas) & Approx. 500mV /5.48 MeV (alphas)
- Noise level: < 50 keV (betas) @ 22 degrees C
- Contamination measurement
- Alpha/Beta descrimination measurements
- ITEM: DATA
- Scintillation material: ZnS(Ag)+plastic scintillator Ø51 x 0.25mm
- Crystal read out: SiPm
- Housing: Black light tight polyester
- Entrance window: Light-tight double aluminized mylar 0.9 mg/cm2
- Protection: 72 % transmission stainless steel grid
- Power supply: + 5 – 16 V (3 mA)
- Electrical connections: LEMO connector ERA0S302CLL Flying leads
- Cable: 50 cm 4 core shielded LIYD11Y
Performance:
- Pulse rise time (0-100%):
- 150 ns (betas)
- 1μs (alphas)
- Pulse fall time (1/e):
- 1.2 μs (betas)
- 3 μs (alphas)
- Gain: Approx. 75 mV/MeV(betas)
- Approx. 500mV /5.48 MeV (alphas)
- Noise level: < 50 keV (betas) @ 22 degrees C