J.U.M. - Model 3-200 -Transportable Fully Heated FID Analyzer
The Model 3-200 is a transportable fully heated FID is successfully used throughout international markets for over 30 years. Also among other portable gas FIDs, the 3-200 It is the first ever built transportable fully heated FID using high pressure back purge sample filter. It is widely distributed throughout the industry and the stack testing community and is much liked for its ruggedness, high accuracy, sensitivity and stability.
Our optional 50 liter highly safe low pressure metal hydride fuel storage can easily be mounted on the left side panel and allows a 45 hour minimum of uninterrupted operation. This fuel storage can be self-refilled at a low pressure of only 25 bar. No special adapter needed. A standard 2-stage pressure regulator with a 30 bar outlet is sufficient
The 3-200 is officially certified to be 100% identical to the VE7 and 3-300A analyzers. TÜV confirmed to fully comply with QAL 1 (EN 14 181 and EN ISO 14956) and EN12619 regulations. They also fully complies with USEPA Method 25A and Method 503.
The 3-200/OVE
is a 100% identical version at 20% lower cost due to the use of a disposable sample filter in the heated oven. With this exception the 3-200/OVE is a 100% identical analyzer.
- Maintenance free sample filter back purge system allows filter to be cleaned without dismantling (automatic optional)
- Built in burner air generator, no external air cylinder needed
- All components in contact with sample fully heated and controlled at 190 degree C
- Built-in sample pressure and sample pumps
- Automatic flame out alarm and contact
- Fast response within 1 second
- Low fuel consumption
- Very selective
- Microprocessor controlled PID-type temperature controller
- Remote control for sample, calibrate and back purge is standard
- Remote or automatic range change optional
- Stack gas hydrocarbon emissions monitoring
- Fence line monitoring
- Raw exhaust vehicle emissions analysis
- Catalytic converter testing
- Measuring engine combustion efficiency
- Hydrocarbon contamination monitoring in air and other gases
- Carbon adsorption regeneration control
- Detection of trace hydrocarbons in purity gases used in the semi conductor industry
- LEL monitor of solvent laden air
Some Options Cannot be Combined:
- AMU 32: Automatic range change
- AP0 32: EXTERNAL automatic programmable backpurge system for the sample filter
- AZM32: Automatic flame ignition and re-ignition
- ENGA 32: 6-digit direct unit display, 0-100.000 ppm
- FOAS 32: Flame out alarm with automatic fuel shut off valve
- ICM 32 *: Built-in NMHC Cutter, measure either THC or Methane-Only with one analyzer
- LT0 32: Measurement of low trace hydrocarbon levels-Requires external, zero grade combustion air supply!
- PDA 32: Sample pressure monitor with alarm
- RCA 32: 0-20 mA analog output instead of 4-20 mA
- RCC32: Remote control range change
- RCIO 32: 0-20 mA analog output, galvanically isolated
- RCI4 32: 4-20 mA analog output, galvanically isolated
- TPR 32 **: Internal temperature controller for heated sample line, e.g. JUMTJ100
- FSS32: Hydrogen fuel storage cartridge including male and female (mounted) 1/4" Swagelok quick connector
- Important!:
- * ICM cannot be combined with LTO
- ** TPR available as external module only when combined with ICM
Availability of options may change without notice!