Umwelttechnik MCZ GmbH

- Model NGA M1000 - Zero Gas System

From

0

Compressed air system with compressor and filter. Oil-free dental compressor. For process gas, lab or dental gas or zero gas system for gas analysis. Modular system for stationary installation. Optionally with catalytic gas purifier and drier.

  • Max. pressure: 8 bar (received working pressure of 2.5 to 4 bar at filter module input)
  • Suction capacity: 95 l/min
  • Pressure tank: 24 l
  • Drying: water separator/adsorption dryer
  • Connected load: 220 V 50 Hz
  • Noise level: 73 dB(A) / 1m
  • Dimensions: 385 x 385 x 570 mm
  • Weight: approx. 57 Kg
  • Other: pressure controller and manometer to set output pressure

Filter Modul:

  • Installation plate with 4 filter cartridges (50 x 450 mm)
  • Dust filter

  • Input pressure: max. 10 bar
  • Output pressure: 2 ... 4 bar, adjustable
  • Max. throughput: 15 l/min at 3 bar
  • Process: catalytic oxidation on heated Pd converter
  • Efficiency: > 99,5 % for BTX/C3Hm, 99,9 % for CO, ethylene
  • Connected load: 220 V 50 Hz / 400 to 600 VA
  • Dimensions: 19 inch / 4RU tabletop case or rack mount
  • Other: pressure controller and manometer to set output pressure, temperature of catalytic gas purifier can be set as high as 500°C, displays catalyst nominal and actual temperature