Metrohm AG
  1. Companies
  2. Metrohm AG
  3. Products
  4. Metrohm - Model Metrosep Carb 2 - ...

MetrohmModel Metrosep Carb 2 - 250/4.0 6.1090.430 -Anion Exchanger Column

SHARE

The Metrosep Carb 2 - 250/4.0 IC column is particularly suitable for the determination of carbohydrates using alkaline eluents and pulsed amperometric detection. The high-capacity anion exchanger column is based on a styrene-divinylbenzene copolymer. It is stable in the range of pH = 0 - 14 and separates monosaccharides and disaccharides. It is also suitable for the analysis of sugar alcohols, anhydrous sugars, amino sugars, etc. The 250 mm version of the Metrosep Carb 2 separation column is optimized for complex separations.

Most popular related searches
  • Carrier material: Polystyrene-divinylbenzene copolymer with quaternary ammonium groups
  • Column dimensions: 250 x 4.0 mm
  • Eluents:
    • Hydroxide/acetate eluent (standard eluent)
      • Sodium hydroxide (c = 20 mol/L): 100 mmol/L; 10 mL/2 L
      • Sodium acetate: 10 mmol/L; 1,640.7 mg/2 L
    • Hydroxide acetate eluent, modified
      • Sodium hydroxide (c = 20 mol/L): 5 mmol/L; 0.5 mL/2 L
      • Sodium acetate: 2 mmol/L; 328.1 mg/2 L
  • Housing material: PEEK
  • Max. pressure: 20 MPa
  • Maximum flow: 0.8 mL/min
  • Organic modifier: In the eluent: 0 - 50% acetonitrile or methanol
  • In the sample: 0 - 100% acetone, acetonitrile, or methanol
  • Particle size: 5.0 µm
  • pH range: 0–14
  • Standard flow: 0.5 mL/min
  • Standard temperature: 30 °C
  • Temperatur range: 20–60 °C