Model cHFM -Compact Hand & Foot Monitor
-
Industry’s smallest hand and foot monitor—with a footprint of less than 2.5 ft²
-
Available options—wall-mount, desk-mount, cart-mount (as shown in the photo)
-
Counting cycles start automatically via photoelectric hand and foot sensors
-
Automatic background compensations—monitors for abrupt change in ambient gamma level
-
Fully plumbed to integrated gas tank or in-house gas supply
-
Inlet and outlet counting gas flowmeters clearly visible for adjustment on the front panel
-
International versions available using SI units and adapted power supply
-
Simple touch screen controls are part of the user-friendly interface
-
Data logs download via the RadNet compliant Ethernet port or the built-in USB port
-
Remote monitoring via the wireless or Ethernet RadNet output option
The electronics assembly provides the user interface and all necessary detector support electronics. It consists of an aluminum dust-tight and splash-proof enclosure containing a Windows-CE 5.0 processor board along with internal counting logic coupled to the detectors via the rear panel mounted SDB detector interface modules. It is removable from the wall mount bracket without tools and uses quick-release cables and gas connectors from the detectors to allow the electronics assembly to be easily detached from the mounting bracket for maintenance.
The front panel includes all indicators and controls. The LCD touchscreen display is used for system configuration as well as to display counting results. Front-panel mounted LEDs are used for status indication; the Alarm
LED (Red) indicates that contamination was detected, the Ready LED (Green) indicates the monitor is ready to perform a count, the Out of Service LED (Yellow) indicates a trouble condition and that the monitor cannot be used. Two Rotameters are also mounted to the front panel to provide inlet gas flow adjustment, and to indicate outlet gas flow. The audible Sonalert™ alarm sounds when contamination is detected and is silenced/reset by pressing the Alarm Acknowledge pushbutton. For counting small articles on the hand detector, a Count pushbutton is provided to manually start a count cycle, thereby overriding the photoelectric sensors.
The power On/Off button is located on the side panel along with the USB port for downloading log data. The click rate speaker driven by the frisker probe is also located on the right side. The rear panel of the electronics assembly includes the 12 VDC charger/ power connector, P-10 gas quick-release connectors, and the Ethernet connector. The detector electronics SDB modules are mounted externally to the rear panel and provide the interface between the data processing computer and the gas proportional detectors. The SDB modules (one per detector) are easily removable without tools.
The wall mount for the hand monitor and electronics assembly is designed so that the LCD display can be positioned at an adjustable height with respect to the hand monitor so that a comfortable hand position is not compromised by the display height. The foot detector assembly can be conveniently placed on the floor or on a cart below the wall mount and allows for an adjustable angle of from 15° to 30° for comfortable foot placement. Both hand and foot monitor assemblies are equipped with 24 cm x 39 cm (9”x15”) gas proportional detectors with MHV connectors for high voltage and signal plus a quick-disconnect connector for the P-10 counting gas. The hand and foot monitors use photoelectric sensors to trigger the count cycle when a user places their hand or foot on the surface of the detector assembly.