Calibration Technologies, Inc.

Model NH3 -Responder Ultra

SHARE
Handheld Ammonia gas detector used for detecting all levels of ammonia for leaks and hazmat response. Detection range from 0-100% LEL of Ammonia (PID: 0-1,000 PPM NH3) (LEL: 4,500 to 150,000 PPM NH3)). Includes Oxygen sensor with range of 0-25% O2 (vol). Smaller and lighter than competitive instruments. Water-resistant design in an integral concussion-proof enclosure. Large LCD allows for easy-to-read monitoring. Integral motorized pump for remote sampling and fast response. Rechargeable lithium polymer battery (14 hours runtime @ -4F). 95 dB alarm tone, bright LED alarm bars, & internal vibrating alarm for noisy areas. Auto backlight: activates in low light & when in alarm. Four alarm levels: instantaneous Low Alarm, High Alarm, TWA and STEL. Records TWA, STEL, and max exposures to gas & displays readings on command. Datalogger with programmable sampling intervals. Available options for additional H2S and CO sensors (call for price and availability). 
Most popular related searches

Size: 5.8″ high x 3.3″ wide x 1.6″ deep

Weight: 14.5 oz (with clip)

Alarms: Visual, Vibrating, Audlble (95 db)

Low, High, STEL, TWA, OL (over limit)

Tests: Activated Detectors automatically perform one internal diagnostic every 24 hours

Pump: Integral motorized pump for remote sampling and fast response

User Options: Adjust STEL period (5-16 minutes in 1 minute intervals), Calibration span levels, Calibration interval, Bump test interval, Select combustibe gases measurement: 0-100% LEL or Methane gas 0-5% LEL and PID correction factors, Calibration due date (1 to 365 days, or set to zero for off), Individual sensor enable/disable, Latching alarms, Stealth mode, Languages: English, French, German, Portugese, Spanish, Simplified Chinese and Russian

Detection Range: PID: 0-1,000 PPM NH3; LEL: 3-100% LEL (4,500-150,000 PPM NH3); O2: 0-25% vol

Features: Calibration Mode, Relay Test Function, Adjustable warn and alarm setpoints, Alarm log, Adjustable relay latch/non-latch, Downscale alarm setting for O2 monitoring, horn silence button, remote reset and silence functionality

Warranty: Instrument: 3 years

PID Sensor: 1 year

LEL Sensor: 3 years

O2 Sensor: 3 years