- Home
- Companies
- PID Analyzers, LLC
- Products
- PID Analyzers - Model 105 - Hydrogen ...
PID Analyzers - Model 105 -Hydrogen (H2) Analyzer with Optional Datalogging
The Model 105 is an analyzer for the measurement of hydrogen (H2) in the workplace. This analyzer is unusual since it is easy to calibrate, has a fast response and has only one step to send logged data to a PC. It uses a thermal conductivity sensor for the measurement of ppm or % levels of hydrogen. It has many unique features. It can also be used as a helium leak detector with a different calibration.
H2 Range 0-2,000 ppm and
0-10,000
Bright OLED Display
Hydrogen-Temperature compensated
Integral pump
Fast response- 90% in 30 seconds Audible alarm
Ease of calibration
Optional Datalogging, Bluetooth
- Worker health & safety
- Monitoring Li Battery production area
- H2 in binary Gas Mixtures
- H2 in N2, H2 in CO2, H2 in Ar, H2 in O2
- Hydrogen leaks in hydrogen storage
Thermal Conductivity (TCD) - Hydrogen analyzer
The Thermal Conductivity Detector (TCD) measures the difference (change in resistance) in cool-ing between the sample and a sealed reference through a Wheatstone Bridge circuit that detects small changes in resistance. This difference in the signal between the measurement and refer-ence sensors is amplified and the output is displayed on the digital meter after processing. The range for the TCD can be 0-20,000 ppm or 0-60 % v/v. The detection limit on the ppm range is approximately 20 ppm.
Options
There are two Models available: H2 (82-105H2-00) or - He (82-105He-00) for leak detection
Optional datalogging includes (storage, Bluetooth, RTC & >100,000 points. Bluetooth (makes it easy to download data to a PC, cloud or a tablet)
- Type: Thermal conductivity sensor for H2 Adjustable audible alarm
- Power: Rechargeable 3.7 V LiPO battery
- Display: 2.4” Bright OLED
- Dimensions: 6 1/8" x 3 5/8" x 1.57”
- Accuracy of Measurements: +/- 2%
- Operating temperature: 0-50 C
- Certifications: CE, RoHS
- Analog channels: 16 bit ADC for best precision & accuracy
- Electrical classification: General purpose, not for hazardous area
- Outputs: RS232, can be used to transfer the limited 1000 points to a PCvia a USB cable
- Datalogging-optional: 100,000 data points includes Bluetooth, RTC, memory; Includesselect BT on to transfer data to PC
- Alarm: Audible, visual (blinking display)
- Sleep function: To extend time w/o battery usage
M 105 Pricing $1,350.00; Optional datalogging, BT RTC $75.00
