PID Analyzers - Model 312 GC -Portable Gas Chromatograph System
PID Analyzers offers a new lightweight portable GC for environmental, quality control, natural gas... applications. The base Model 312 comes with syringe injection, oven, power supplies & Pentium PC with Windows XP (embedded) and PeakWorks chromatography software in a rugged plastic case.
Most popular related searches
portable gas chromatography
portable gas chromatography system
catalytic combustion detector
thermal conductivity detector
flame ionization detector
gas chromatography system
flame detection system
gas valve
combustion detector
photometric detector
- PID Lamps: Our 9.5 and 10.6 eV lamps remain the best in the business since we developed them over 30 years ago.
- Display: Our 7`` Color VGA display (640 x 480 resolution) is easy to read. The display is a touch screen.
- ADC: Our 16 bit ADCs provide a resolution of 1 part in 64,000 and our signal algorithm minimizes noise by signal averaging in the msec range; the ADC can acquire positive or negative signals.
- Digital outputs: USB connection.
- Oven:
- Size:
- Typical column diameter: 3.5``
- Temperature range:
- Battery: 25-50 Degrees Celcius
- External power: 25-100 Degrees Celcius
- FID: Our very sensitive FID has flame out sensor that will shut off the hydrogen if the flame goes out.
- PC: Embedded Pentium Computer with > 1 GByte of RAM; data storage on 1 Gbyte USB memory stick.
- Autoranging: The 312 has 3 ranges with computer controlled autoranging- x1, x 10, x 100.
- Battery life: 3-4 hours; can operate off cigarette lighter adapter.
- FPD: Our very sensitive FPD has flame out sensor that will shut off the hydrogen if the flame goes out.
- Operating system: Windows XP® embedded stored on 2.5 GByte smart card (the PeakWorks program and all data are on the memory stick.
- Autozero: Beginning of each run.
- Carrier gas: 50 liters at 1500 PSI; typical capillary column 15-20 cc/min; typical packed column 30-40 cc/min.
- Ease of calibration: Manual.
- Peak Detection & System Operation: Controlled by our 32 bit PeakWorks software program with more than 20 years of development.
- Programmable Voltage/current output: 0-1VDC.
- Programmable setpoints: The low and high setpoints (depends on the options) can be customer programmed for each point and/or compound.
- Detectors: Photoionization detector (PID), Flame ionization detector (FID), Thermal conductivity detector (TCD), Flame photometric detector (FPD), Far UV absorbance (FUV) detector, Catalytic combustion detector (CCD)
- Linear Dynamic Range: PID - 108 FID - 2 x 106 TCD - 2 x 104 FPD- S - 103; P 105 Far UV - 106 CCD - 5 x 104
- Sample Flow Rate: > 0.5 LPM
- Power consumption: < 50 watts maximum; 115 VAC, 0.5 amps (12 VDC 5 amps)
- Species measured:
- PID & FID: Organics (VOCs by PID or FID) and inorganic species (H2S, NH3, AsH3… ) that can be ionized by the PID lamp (9.5, 10.6 or 11.7 eV)
- TCD: Universal detector- fixed gases (O2, N2, CO…), organic and inorganic compounds, very sensitive for H2, He, Xe
- FPD: Selective response for S or P compounds
- Far UV: Nearly universal response for organic and inorganic compounds
- CCD: Combustible compounds such as hydrocarbons
- Sensor(s): 1 or 2 detectors
- Precision: +/- 1-2% of reading
- Injection: On-column syringe injection or automatic with 6 port gas valve (optional)
- Span Drift: <1% over 24 hours
- Dimensions: Enclosure - 18.75``x 13.75`` x 7.5``
- Weight: Fiberglass case - 27.5 pounds
- Detection Limit:
- PID - 0.5 ppb benzene
- FID - 50 ppb benzene
- TCD - H2- 10 ppm; O2 150 ppm
- Response time: Depends on retention times
- Zero Drift: NA- autozero on each run
- Instrument operating conditions: 5-40 degrees C, 0-95% RH (non-condensing)