- Home
- Companies
- Vacuum Metrology Inc.
- Products
- Vacuum Metrology Inc - Model VMS2Ph - ...
Vacuum Metrology Inc - Model VMS2Ph -Portable Calibration System for Lyophilization Gauges
The VMS2Ph is a high-accuracy calibration system tailored for the pharmaceutical industry, specifically for the calibration of lyophilization gauges. Developed by Vacuum Metrology Systems, this portable unit is compliant with CFR Title 21, Part 11, offering automatic, semi-automatic, and manual modes. It comprises a robust Calibration Management Software that ensures the system meets ISO 9001, ISO/IEC 17025, and ANSI Z540 standards. It features up to three high-accuracy NIST or PTB traceable vacuum transfer standards, with a basic accuracy of up to ±0.025%. Its pressure control capabilities range from 1.3x10^-1 to 1.3x10^4 Pa, maintaining a resolution of 0.01% of the set point, enabled by a proprietary algorithm. The system also includes data security measures that provide secure databases to maintain a clear audit trail enhancing the VMS2Ph's reliability as a precise, traceable solution for pharmaceutical vacuum gauge calibration.VMS2Ph (Pharmaceutical) is a High Accuracy model from a new family of calibration systems developed by Vacuum Metrology Systems. VMS2Ph system is specifically configured to fit the requirements for calibration of Lyophilization gauges. The system has built in technical controls for compliance with CFR Title 21, Part 11.
VMS2Ph is a truly portable fully automatic calibration system for onsite calibration of vacuum gauges for pharmaceutical industry. In addition to portability and unsurpassed accuracy, VMS2Ph offers a built in Calibration Management Software suite to ensure compliance with ISO 90XX, ISO/IEC 17025, ANSI Z540 and other industry specific quality standards.
The system utilizes up to 3 high accuracy NIST or PTB traceable vacuum transfer standard with basic accuracy of up to .025% . The standards coverage can be custom configured to provide the best accuracy throughout the required calibration pressure range. The proprietary pressure control algorithm allowins for the repeatable control resolution of up to 0.01% of the set point.
VMS2Ph system is fully compatible with the requirements of ISO-3567.
VMS2Ph can be configured to utilize the Transfer Standards that your company already owns.
High Accuracy Transfer Standards
up to 3, NIST or PTB traceable, with basic accuracy of up to ±0.025% of the reading
Number of Test Ports
3
Base Pressure
Less than 7x10-5 Pa (5x10-7 Torr)
Pressure Control Range
Fully Automatic Pressure Control from 1.3x10-1 to 1.3x10+4 Pa (1x10-3 to 1x10+2 Torr)
Pressure Control Resolution
Up to 0.01% of set point
Operating Modes
Automatic, Semi-Automatic, Manual, Service
Data Averaging
10 to 99 points
Thermal Transpiration Correction
On/Off for Standards and/or UUTs
(Calculations per Takaishi-Sensui formula)
Process/Calibration Gas
Nitrogen or Dry Air
Dry Pumping System
Scroll + Turbomolecular pumps
Fully Enclosed Cabinet
Eliminates temperature variations and removes the uncertainty contributions caused by the temperature drift.
Cleanroom Compatibility
Fully Compatible
ISO Compliance
Fully compliant with ISO 3567: Vacuum Gauges — Calibration by direct comparison with a reference gauge
Data and System Security
Secure databases for clear audit trail:
- process recipes
- calibration data
- Event logger database
- Technical controls for compliance with CFR Title 21, Part 11
Calibration Management Software Suite
- Create and Print Calibration Certificates
- View and Track calibration history and Trends
- Create and Save unlimited Calibration and UUT recipes
Power Requirements
110/230 VAC, 50/60 Hz, 15A