ENVEA - Model NDIR-GFC - MIR 9000P -Portable Multi-Gas Analyzer
The MIR 9000P belongs to the new generation of ENVEA gas monitors, showcasing eco-design, IoT and on board intelligence. It measures simultaneously 8 gases – NOx, SO2, CO, CO2, CH4, N2O, O2 and residual H2O – and uses the non-dispersive infrared method (NDIR-GFC) with gas filter correlation. O2 is measured by a built-in SRM paramagnetic sensor. The analyzer has been designed to meet the specific needs of on-site regulatory measurement providing superior mobility, robustness, accuracy and compliance. All this makes up for a unique gas emission monitor offering high productivity and low operational cost.
-
INCREASED MOBILITY & PRODUCTIVITY
- Lightweight (15 kg/33 lbs), compact and extra resistant transport case
- Standby mode guarantees the instrument is operational at any time
ROBUST DESIGN
- Extended operating temperature ranges
- High protection (IP 44) against water splashing from any direction
STAY CONNECTED & GET ON WITH OTHER TASKS
- Remote access to full operation thanks to ENVEA Connect™ App and proprietary WiFi
- Smartphone alerts and notifications
ONE CLICK OPERATIONS
- Integrated user guidance and shortcuts to key functions: calibrations, span check, zero, etc.
- Secured user management
QUICK INSTALLATION
- Warm-up time: 30 min maximum
- Robust gas connectors
- Self-adaptive measuring scales
- Compatible with most sampling systems on the market
PREPARE REPORT ELEMENTS WHILE ON SITE
- Direct access to real-time and stored data ; time frame selection to generate graphs
- All data can be saved on USB memory stick
MAXIMIZED UPTIME OF THE ANALYZER
- Turn-up lid designed for quick access to operational modules and easy maintenance
- Modular design and Plug & Play parts (O2 probe, NOx converter, protection filter, etc.)
- Regulatory and periodic stack emission monitoring for industrial combustion sites, boilers, etc.
Measuring Ranges
- NOx as NO (after NOx converter): 0-70/2000/3300 mg/Nm3 0-50/1500/2500 ppm
- NOx as NO2 (after NOx converter): 0-110/3000/5000 mg/Nm3
- SO2: 0-141/1410/8500 mg/ Nm3 , 0-50/500/3000 ppm
- CO: 0-60/3000/8000 mg/Nm3, 0-50/2400/6500 ppm
- CO2: 0-20/30% vol.
- N2O: 0-1000 mg/Nm3, 0-500 ppm
- CH4: 0-70/700 mg/Nm3, 0-100/1000 ppm (higher range upon request)
- O2: 0-10/25% vol. measured by a built-in paramagnetic sensor Residual H2O: 0-2% vol.
Technical Specifications
- QAL1 to EN 15267-4 certified: Pending
- US EPA compliance: YES
- Sampling technology: Cold/Dry Extraction (gas cooler)
- Sample transfert: non heated polytube / heated sampling line
- back flush function: N/A
- span injection: YES
- Built-In O2 option: YES
- Serial link communication: 3 x USB ports type A: 2.0 (compatible 3.0)
- 1 x RJ45 (MODBUS TCP, MODBUS RTU (dongle) UDP Protocol)
- Other Communication: WiFi (via USB)
- Analogue OUTPUTS: Option: 8 analog outputs (0-1 V, 0-10 V, 0-20 mA, ou 4-20 mA)
- Analogue INPUTS: Dynamic 250 mV or 10V (only one input can be used at once)
- Dimensions (mm): 641 x 209 x 393mm (WxDxH)
- QAL1 certification by the TÜV according to EN 15267-1, EN 15267-2 and EN 15267-4. Download the TÜV certificate: 0000074636
- Compliant with EN 15267-4
Air quality – Certification of automated measuring systems – Part 4: Performance criteria and test procedures for automated measuring systems for periodic measurements of emissions from stationary sources - Compliant with US EPA