Nova Analytical Systems Inc. - part of the Tenova Group

NOVAModel 600 Series -Portable Multigas Ambient Air Analyzer

SHARE

For ambient air or confined space monitoring for up to any four choices of O2, CO2, NO, NO2, CO, LEL combustibles, temperature, and % RH. Single gas versions are also available for H2S, SO2, Cl2, HCL & NH3.

Most popular related searches
  • Method of Detection : O2, CO, SO2, NO, NO2, H2, CL2, NH3 and H2S by electrochemical sensor, % RH by dielectric ceramic sensor, CO2 by infrared sensor, combustibles by catalytic sensor
  • Ranges Available : 0-100 PPM NO0-100% RH0-20 PPM CL20-20 PPM H2S0-20 PPM NH30-20 PPM NO20-200 PPM CO0-200 PPM H20-25.0% O20-50 PPM SO20-50°C Temp.0-5000 PPM CO2
  • Resolution : 0.1% on % gas, 1 PPM on PPM gas
  • Accuracy & Repeatability : ±2% of full scale
  • Drift : Less than 1% of full scale or 1 PPM per 8 hours, whichever is greater, with constant ambient temperature
  • Response Time (T-90) : Varies with individual sensor but generally under 2 minutes for full reading
  • Ambient Temperature Range : 4 to 49 ºC40 to 120 ºF
  • Linearity : ±2% of full scale or 1 PPM, whichever is greater
  • Size & Weight : 15" W x 7" H x 12" D @ 14 lbs. (374 x 175 x 305 mm @ 6.4 kg)
  • Power : 115 VAC, 60 Hz for recharging220 VAC, 50 Hz
  • Output Options : 0-1 VDC4-20 mA
  • Alarms : Audible and visual alarms are optionally available as well as low battery and low flow
  • Gases Measured : 0) NH3 - 0-20 PPM1) O2 - 0-25.0%2) CO - 0-200 PPM3) CO2 - 0-5000 PPM4) LEL Combustibles (not avail. for 501 Series)5) H2S - 0-20 PPM6) SO2 - 0-50 PPM7) NO - 0-100 PPM8) NO2 - 0-20 PPM9) CL2 - 0-20 PPM

For ambient air or confined space monitoring for up to any four choices of O2, CO2, SO2, NO, NO2, H2, CO, Cl2, NH3, H2S, LEL combustibles, temperature, and % RH.

  • Rugged design that is easy to operate and maintain
  • Fast warm-up and response
  • Separate electrochemical sensor for O2, CO, SO2, NO, NO2, H2, CL2, NH3, and H2S
  • Sensitive infrared detector for CO2
  • Catalytic sensor for combustibles
  • Capacitive ceramic sensor for % RH
  • Digital readout with backlighting for each gas
  • Built in sampling pump, flow meter, filter and 12 ft. sampling hose
  • Rechargeable battery
  • Rugged suitcase (K) style case
  • Recorder outputs of 0-1 V or 4-20 mA
  • Audible and visual alarms
  • Built in printer
  • Detachable/portable data logger
  • Nitrogen (N2) for zero
  • Analyzed gas cylinders for span