SHARE

The Model 113+ is a EC based Analyzer that can measure up to ppm to% v/v of components and has the capability to add up to 2 EC sensors and a CG or TCD sensor. The Model 113 includes the choice of one Snap-On Module plus up to 1-3 optional electrochemical sensors an an optional TCD or CG sensor , datalogging, Operator`s Manual, and Rechargeable Nickel Metal Hydride Battery.

Most popular related searches
  • Snap on Heads: Unique Snap on Heads are available only from PID. The readout unit recognizes the head type and calibration factors when the heads are interchanged.
  • Datalogging: Manual or automatic datalogging (7,000 points) is available. The 113 has a built in real time clock. Viewer software (Windows 98-XP compatible) comes with the 113.
  • Sensors: EC (approximately 30 optional EC sensors), TCD, or CG.
  • Library: There are libraries for the TCD & CG built into the 113
  • ADC: Our 16 bit ADC provides a resolution of 1 part in 64,000 and our signal algorithm minimizes noise by signal averaging in the msec range
  • Single piece construction: The single piece construction (when the head and readout units are snapped together) of ABS plastic is very rugged and durable
  • Display: Our 2 line by 16 character display can be backlit. The upper display provides the concentration and the bottom display provides the sensor, units, alarm and logging mode.
  • Ease of calibration: The two point calibration for the TCD uses a zero gas and a span gas. The time to calibrate is about 40 seconds. The Cal button is located on the main menu for ease of use.
  • Backlit Display: Allows for easy visibility.
  • Programmable Voltage output: 0-1VDC
  • Species measured: Universal response for TCD
  • Precision: +/- 2-3% of reading
  • Range: ppm to 100% v/v depends on the sensor
  • Weight: 1.5 pounds
  • Sensor(s): CO, O2, O3, H2O2, SO2, H2S, THT, NO, NO2, NH3, AsH3, PH3, SH4, HCl, Cl2, ClO2, F2, HBr, HF, HCN, I2, Phosgene, H2, ETO, Formaldehyde, C2H4, 1,3 BD, Acrylonitrile
  • Response time: <10 sec. to 90% response plus … depending on the sensor
  • Linear Range: ppm to 100% direct reading
  • Instrument operating conditions: 5-40 degrees C, 0-95% RH (non-condensing)
  • CG Sensor: Hydrocarbons, H2, NH3 …
  • Sample flow rate: > 0.2 LPM
  • Power consumption: < 0.75 watt maximum
  • Performance: Isobutylene Referred
  • TCD Sensor: H2, He, Xe, CO2, …
  • Span Drift: <3% over 24 hours
  • Dimensions: 8`` L x 3`` W x 2.25`` D
  • Battery operated: > 10 hours of operation