- Sonation - Model PRSO-V2 - Column Oven
Sonation - Model PRSO-V2 -Column Oven
Developed at the Max-Planck-Institute of Biochemistry in Martinsried, the Sonation column oven is used to regulate the temperature of the separation column in ESI-Nanospray applications. Constant column temperature makes measurements significantly more accurate, and higher column temperatures mean that significantly higher flows can be achieved while maintaining the same pressure. As a result, the measurement and wash cycles can be significantly accelerated and the utilization of the instrument is improved.
The Sonation column oven has been developed with the goal to solve three problems in the ESI-Nanospray workflow:
The influence of a changing room temperature.
In most labs the room temperature is not constant. Air conditions or opened doors are causing a changing temperature that influences the results.
The length of the column is limited due to the maximum pressure of the used HPLC.
The longer the used column, the higher the pressure must be to maintain the same flowrate.
The washout of the columns consumes a lot of time.
Even with the maximum pressure of the HPLC the washing cycles needs a lot of time.
With the Sonation column oven the temperature of the column is now completely independent from the room temperature. The column can also be heated to high temperatures to accelerate the washout cycles or getting higher flowrates during the run. Also the use of longer columns is now possible.
With the software that comes with the oven the temperature can be adjusted direct on the PC. It can also be switched between two temperatures automatically using an instrument method in Xcalibur.
Hardware specifications:
- Usable column lengths: From 13 cm, no limit in maximum length
- Usable column diamters: Up to 2mm
- Temperature control range: 15-60 °C software controlled
- Control accuracy: >0.1°C
- Available for the following sources:
- Thermo Sientific ES071
- Thermo Sientific ES072
- Bruker Captive Spray
- Operation with other sources is possible, but requires mechanical adjustment. If you have another source in use, please contact Sonation.