Teledyne API

- Model T108 - Total Sulfur Analyzer

From

0

The Model T108 Total Sulfur analyzer is designed to measure mixed sulfur impurities, collectively referred to as Total Sulfides, in air or carbon dioxide gas. Since there is no scrubber in the system, the instrument reading is the sum of the oxidized sulfur compounds and SO2.

The Model T108 consists of a modified Model T100 SO2 analyzer with special software and a Model 501TS high temperature thermal oxidizer. Sulfur compounds are heated as they pass through the converter and oxidized into SO2. When analyzing CO2, which generally contains no oxygen, approximately 6% O2 is added to the sample before entering the converter. This dilution is corrected in the calibration procedure. All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/0, and built-in data acquisition capability. All instrument set up, control and access to stored data and diagnostic information is available through the front panel, or via RS232, Ethernet, or USB com ports, either locally or by remote connection.

The Model T108 comes with NumaView™ Software. NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView™ Software.

  • Ranges: 0-5 ppb to 0-20,000 ppb, user selectable
  • Dual ranges and auto ranging
  • Large, vivid, and durable color graphics display with touch screen interface
  • Ethernet, RS-232, and (optional) USB com ports
  • Front panel USB connections for peripheral devices and firmware upgrades
  • 8 analog inputs (optional)
  • Adaptive signal filtering optimizes response time
  • Temperature & pressure compensation
  • Continuous Auto Zero Correction
  • Comprehensive internal data logging with programmable averaging periods
  • Ability to log virtually any operating parameter
  • Two-year warranty​

  • Ranges :
    • Min: 0-50 ppb full scale
    • Max: 0-20,000 ppb full scale (selectable, dual range supported)
  • Measurement Units  : ppb, ppm, μg/m3, mg/m3 (selectable)
  • Zero Noise : 50 ppt (RMS)
  • Span Noise : 0.5% of reading (RMS) above 5 ppb
  • Lower Detectable Limit  : 0.4 ppb
  • Zero Drift : < 0.5 ppb/24 hours
  • Span Drift : < 0.5% of full scale/24 hours
  • Lag Time : 20 seconds
  • Rise/Fall Time  : < 100 seconds to 95%
  • Linearity : 1% of full scale
  • Precision  : 0.5% of reading above 50 ppb
  • Sample Flow Rate  : 650 cc/min ±10%
  • Power Requirements  : 100V-120V, 220V-240V, 50/60 Hz
  • Analog Output Ranges  : 10V, 5V, 1V, 0.1V (selectable)
  • Recorder Offset  : ±10%
  • Included I/O  :
    • 1 x Ethernet: 10/100Base-T
    • 2 x RS232 (300-115,200 baud)
    • 2 x USB device ports
    • 8 x opto-isolated digital outputs
    • 6 x opto-isolated digital inputs
    • 4 x analog outputs
  • Optional I/O  :
    • 1 x USB com port
    • 1 x RS485
    • 8 x analog inputs (0-10V, 12-bit)
    • 4 x digital alarm outputs
    • Multidrop RS232
    • 3 x 4 - 20mA current outputs
  • Operating Temperature Range  : 5 - 40ºC (with US EPA Equivalency)
  • Dimensions (HxWxD)  : Analyzer 7” x 17” x 23.5” (178 x 432 x 597 mm)
  • Converter: 7” x 17” x 22” (178 x 432 x 559)
  • Weight
    • Analyzer: 45 lbs (20.5 kg)
    • Converter: 24 lbs (10.8 kg)

Customer reviews