- Home
- Companies
- WTW - a xylem brand
- Products
- WTW - Model CR 3200 - WTW -1P22 - ...
WTW - Model CR 3200 - WTW -1P22 -Thermoreactor
Digestions of 24 round cells with pre-programmed and user-defined programs. In total 24 round cells (16 mm) in two seperate heating blocks using the same progra.
- 8 stored programs for easy and fast operation including COD fast digestion
- 8 user-defined programs from room temperature/25-170°C with high temperature digestion
- AQA/IQC with optional external temperature sensor for documentation
Programs for standard parameters COD including fast digestion (148°C, 20 min.), total nitrogen, total phosphorus, TOC and heavy metals of galvanizing with automatic heating and switch-off quick. User defined programs can be easily entered and stored from room temperature respectively 25-170°C to perform special digestions and thus allow high temperature digestion.
The very same program can be run for all vials at the same time; to perform two parallel digestion programs, the CR 4200 is the suitable choice. For the analytical quality assurance (AQA), the external temperature sensor TFK CR can be connected to the PC/printer instead of a round cell.
Programs for standard parameters
- 1148 °C for 120 minutes
- 2120 °C for 30 minutes
- 3120 °C for 60 minutes
- 4120 °C for 120 minutes
- 5100 °C for 60 minutes
- 6148 °C for 20 minutes
- 7150 °C for 120 minutes
- 8100 °C for 30 minutes
- Reactor type: Dry temperature control device with safety cover
- Cell shafts: 2 x 12 cell shafts for reaction cells, diameter: 16 ± 0.2 mm
- Reaction time setting: program settings or user-defined programs up to 180 min
- Temperature setting: stored program, user-defined settings from room temperature up to 170°C
- Controlling accuracy: ± 1 °C ± 1 digit
- Temperature stability: ± 0.5 K
- Overtemperature protection: 190 °C ± 5 °C
- Dimensions: D x W x H: 312 x 255 x 185 mm
- Weight:: 4 kg
- Output: unidirectional RS232 interface with 2 banana sockets for external temperature sensor (article-# 250 100), printer, PC
- Power supply: 115/230 VAC 50/60 Hz switchable via fuse setting
- allowed voltage tolerance: ±15 % maximum power consumption: 560 W, Fuse 2x6.3 AT
- Protection Class: IP 20 according to EN 60529
- Ambient temperature: Operation +5 °C to +40 °C
- Climatic class: 2 according to VDI/VDE 3540
- Relative humidity: Yearly mean: < 75 %
- 30 days /year: 95 % Other days: 85 % Light dew: yes
- Site altitude: max. 5000 m above sea level
- EMC: EN 61326-1
- FCC 47 CFR Part 15
- Test certificates: cETLus, CE
- Safety standards: EN 61010-1
- EN 61010-2-010
- UL 61010-1
- CAN/CSA C22.2#61010-1
- CAN/CSA C22.2#61010-2-010