KanomaxModel ANSI / ASHRAE 110 - DIF-KIT -Tracer Gas Diffuser Kit

SHARE

The Kanomax tracer gas diffuser is used in performing tracer gas tests in accordance with ANSI / ASHRAE Standard 110. The diffuser is designed to the specifications of standard drawing #110-83M. The diffuser is placed in the fume hood and sulfur hexafluoride gas is injected at at supply pressure of 30 psig. The internal critical orifice ensures a flow rate of 4 liters per minute. Other orifice sizes can be inserted.

Most popular related searches

The ANSI/ASHRAE 110 test is a method of testing the performance of laboratory fume hoods. The Kanomax Dif-Kit is ideal for use in performing the Tracer Gas test in accordance with the ANSI/ASHRAE Standard 110-1995. Kanomax also offers alternative tracer gas systems due to concerns to prevent greenhouse gas emissions.

  • Diffuser is made to the design and specifications of ASHRAE Standard Drawing #110-83M
  • The internal critical orifice ensures a flow rate of 4 liters per minute (lpm)
  • Thermal anemometer for face velocity testing
  • Commercial fog machine for flow visualization
  • Fume hood performance test according to ASHRAE Standard 110-2016

Dif-Kit General Specifications

  • Critical Orifice: 4 Liters per Minute
  • Regulator: Dual Stage, Specialty Gas
  • Flow Meter: Calibrated 150mm/200psi, Glass Tube
  • Pressure Gage: Large Dial, 0 to 60psi
  • Warranty: 1 year