TrackSense - Model Pro XL -Data Logger for Pressure and Rigid Temperature Sensor
The combined Pressure and Rigid Temperature Sensor has a pressure range of 15 mBar to 6 Bar Absolute, with an accuracy of ±0.25% Full scale, and a temperature range of -20 to +150 °C with a measuring accuracy of 0.05 °C. The sensor can be used for a wide range of applications within the food, pharmaceutical and medical industries. The high quality wireless TrackSense Pro XL data loggers consist of high resistant stainless steel and offers high-end technology that allows for stable and highly accurate measurements within different thermal processes.
The TrackSense Pro data logger series are equipped with an Interchangeable sensor option, unlocking several measuring parameters:
- Temperature
- Pressure
- Conductivity
- Relative humidity
- CO2
- Vacuum
- Steam Penetration
The data loggers can easily be activated and read by the Multi reader station and combined with the Sky module for real time data via wireless communication.
- One logger - multiple sensor options
- 21 CFR Part 11 Compliant
- Real time data with the SKY module
- User-friendly validation software
- Usable with E-Val Pro thermocouple system
Sensor with this logger configuration: Pressure and Rigid Temperature Sensor
- Pressure Measuring Range: 15 mBar to 6 Bar ABS (Calibrated 0 to +140 °C)
- Measuring Principle: Piezoresistive
- Sensor Element: Strain Gauge
- Accuracy: ± 0.25% Full Scale
- Temperature Measuring Range: -20 to +150 °C (Calibrated 0 to +140°C)
- Temperature Accuracy:
- T-63%: 0.8 Second
- T-90%: 1.7 Second
Logger with this sensor configuration: Pro XL 3G
- Operating Pressure: 0.001 mBar to 10 Bar ABS
- Operating Temperature: -20 to +150 °C
- House Material: 316 L Stainless Steel
- Diameter: 25 mm
- Length: 68.8 mm
- Weight With Battery: 80 Grams
- Memory Capacity: 120,000 Data Points / 40,000 Samples
- Minimum Sample Rate: 1 Second
- Maximum Sample Rate: 24 Hours
- Maximum Start Delay: 14 Days
- Intrinsically Safe: Ex II1GD Ex ia IIC T3 Ga, -55°C ≤ Tamb ≤ +105°C
- Time Accuracy: ± 5 Seconds Per 24 Hours
- Battery: TSP 150L Battery
Please Note: If equipment is used in ATEX environment, special conditions for safe use stated in ATEX certificates section 17 must be considered.