WTW - a xylem brand
ProfiLine - Model pH/Cond 3320 - 2EA310 -Universal Multi-Parameter Portable Meter
Universal multi-parameter portable meter with two inputs for pH, mV, ISE and conductivity measurements for demanding applications.
Most popular related searches
multiparameter portable meter
multiparameter meter
life science
data logger
conductivity monitoring
dissolved oxygen meter
conductivity resistivity
conductivity/TDS
waterproofing
TDS
- Two inputs for the simultaneous measurement of pH/mV/ISE and conductivity
- Backlit graphic display for the parallel display of the measured values
- Perfect for monitoring process applications
The pH/Cond 3320 has a robust, waterproof housing (IP67) and is equipped with a memory as well as a data logger with an adjustable interval between 1 s and 60 minutes. All entries contain GLP-supporting data, like the date, time and ID number. The modern, waterproof USB interface is used for easy and quick data transfers. It can be used almost anywhere - in process chemistry, Life Science, food and beverage industry all the way to the pharmaceutical industry (measurement of pH and conductivity according to pharmacopeia).
- Parameter: pH/mV/Temp/ISE
- Measuring range pH: -2.000 ... 19.999
- Measuring range Temperature: -5.0 ... 105.0 oC
- Measuring range mV: -1250.0 ...+1250.0 / -2500...+2500 mV
- ION/Concentration: mg/l mmol ppm %
- Accuracy pH: +/- 0.005/ +/-0.01/0.1
- Accuracy mV: +/- 0.3 mV/ +/- 1 mV
- Accuracy Temperature: +/-0.1 K
- Parameter: Concentration/Saturation/Partial pressure/Temperature
- Concentration: 0.00 … 90 mg/l
- Saturation: 0.0 ... 600 %
- Partial pressure: 0.0 ... 1250 hPa
- Temperature: -5.0 ... 50.0 °C
- Resolution Concentration: 0.01 mg/l
- Resolution Saturation: 0.1 %
- Resolution Partial pressure: 0.1 hPa
- Accuracy Temperature +/- 1 digit: +/-0.1 K
- Accuracy Messwerte +/- 1 digit: +/- 0.5 %
- Parameter: Conductivity/Salinity/spec. resistivity, TDS
- Conductivity: 0.001 µS/cm ... 1000 mS
- spezischer Widerstand: 0.00 ... 199 MOhm cm
- Salinity: 0.0 ..70.0
- TDS: 0 ... 199.9 g/l
- Temperature: -5.0 ... 105 °C
- Resolution Conductivity: 0.1 µS/cm
- Salinity: 0.1
- Accuracy Temperature +/- 1 digit: 0.1 K
- Accuracy Messwerte +/- 1 digit: +/- 0.5 %