Waltron - Model 4000 LDO -Single Channel Analyzer
The Waltron 4000 DO Analyzer offers precise, user-friendly monitoring of dissolved oxygen. Compatible with the Waltron 9165S Smart Sensor, it has a wide measurement range for diverse water applications. This durable device supports flexible installation and multiple output options, ensuring top-quality water analysis in any setting.
Most popular related searches
luminescent dissolved oxygen analyzer
luminescent dissolved oxygen
dissolved oxygen analyzer
single channel analyzer
sample temperature
water analysis
oxygen analyzer
dissolved oxygen monitoring
industrial smart sensor
water quality analysis
- Single-Channel transmitter pairs seamlessly with Waltron’s industry-leading 9165S Smart Sensor
- Easy to configure, capable of displaying dissolved oxygen as ppb, ppm, or %O2
- Durable 1/2-DIN IP66 NEMA 4X enclosure can be panel, surface, or pipe-mounted
- Simple, intuitive menu, with soft-touch button controls
- Two relays and two available 4-20mA outputs
- Large, clear LCD backlit display
- Pairs with the Waltron 9165S LDO Smart Sensor for multiple ranges: 0 to 200ppb, 0-2000ppb, and 0-20ppm (sensor-defined)
- Choice of 12-30VDC or 90-265VAC versions
- Parameter: Dissolved Oxygen
- Technology: Luminescent (Optical)
- Range(s): 0-200 ppb, 0-2000 ppb, 0-20 ppm, 0-45ppm (sensor dependent)
- Accuracy: ±0.5ppb at 1ppb or ±2% of displayed value (whichever is greater)
- Sample Streams: 1
- Response Time: t90 < 30 sec
- Sample Frequency: Continuous
- Analog Outputs: Single current output, selectable 0-20mA or 4-20mA into 750 ohms max, fully isolated to 2kV. Expandable across the full operating range and offset anywhere in that range. Loop fault detection.
- Alarms: 2 normally open, fully configurable setpoints, with volt-free contacts for each relay.
- Calibration Method: Bias/Zero & Sensor Exchange
- Ambient Conditions: 0-55°C (32-131°F)
- Sample Temperature: 32-158°F / 0-70°C; 265°F / 130°C high-temperature CIP version available
- Sample Flow/Pressure: Max 145psig (10bar)