- Home
- Companies
- Hangzhou Chunlai Technology CO., Ltd.
- Products
- Zetian - Model OGA-100 - Odor Gas ...
Zetian - Model OGA-100 -Odor Gas Analyzer
The OGA-100 Odor Gas Analyzer is an independently developed gas analysis product designed for online monitoring of odor gases from unorganized emission sources and main discharge outlets. Based on characteristic absorption spectroscopy and the PLS algorithm, it can simultaneously monitor various odor gases, including methyl mercaptan (CH4S), dimethyl sulfide (C2H6S), dimethyl disulfide (C2H6S2), styrene (C8H8), carbon disulfide (CS2), hydrogen sulfide (H2S), ammonia (NH3), trimethylamine (C3H9N), TVOC, and odor concentration (OU value). Additionally, other components such as SO2, NO, and NO2 can be included according to user requirements.
Most popular related searches
odor analyzer
air quality monitoring system
gas analyzer
industrial air quality monitoring
air quality monitoring
emissions monitoring
high-resolution spectrometer
calibration gas zero
gas analysis device
industry air quality
- Capable of directly monitoring the concentration of various odor gas components with fast response speed
- Uses a high-resolution spectrometer with a low detection limit
- PLS algorithm effectively eliminates cross-interference between components
- Utilizes a long-life pulsed xenon lamp as a light source, with no optical moving parts in the device
- High instrument stability
- Built-in constant temperature components, minimal temperature drift, and strong environmental adaptability
- Atmospheric odor unorganized emission monitoring
- Residential area ambient air odor monitoring
- Mobile vehicle odor gas monitoring
- Chemical industrial park unorganized emission monitoring
- Waste treatment plant unorganized emission monitoring
- Principle: C2H6S, C2H6S2 , C3H9N, C8H8 , CH4S, CS2 , H2S, NH3 : characteristic absorption spectroscopy TVOC: PID; OU: empirical formula inversion
- Components & ranges: C2H6S, C2H6S2 , C3H9N, CH4S, H2S, NH3 , TVOC (optional): 0~5μmol/mol; C8H8 , CS2 : 0~1μmol/mol; OU: 0~1000
- Repeatability: Within 3%
- Stability: ≤5%
- Zero drift: Within ±1%F.S./24d
- Span drift: Within ±3%F.S./24d
- Response time (T90): ≤120s
- Digital signal: 13 outputs, 6 inputs, configurable
- Protocol: HJ 212-2017, Modbus RTU, Modbus TCP
- Comm: RS232, RS485, RJ45
- Sample flow rate: 2L/min
- Calibration gas: Zero gas: Dry nitrogen or clean air
- Span gas: Must be within 90%–110% of the target range; unsuitable if outside this limit
- Ambient temp: (0~40)℃
- Ambient humidity: ≤85%RH, non-condensing
- Power supply: AC (220±22)V, (50±1)Hz
- Power consumption: About 150W
- Dimensions & weight: 483(W)*530(D)*221(H)mm (excluding protruding parts), about 30kg