In order to keep your Environics Series 7000 Zero Air Generator (ZAG) working at the highest level, we recommend the following maintenance schedule, assuming average system conditions and use. Fan filter should be cleaned by blowing out the filter monthly. The pollution scrubber/converter and secondary scrubber media should be replaced yearly. The input and output particle filters should be replaced yearly. Depending on use, the pump seals may require replacing every 2-4 years (or longer). If your older ZAG pump runs continually, this is typically the cause. As a reminder, it is important that the drain be free of obstructions.  Connect a line to the drain port and route to a water collection container.

Most popular related searches

Keep running at peak performance

We are happy to offer replacement parts to help you keep your Environics ZAG in tip-top shape. Please complete the brief form below to request parts. We will contact you to finalize your order.