2B Technologies
  1. Companies
  2. 2B Technologies
  3. Products
  4. 2B-Technologies - Model OEM-106-L, -M, ...

2B-TechnologiesModel OEM-106-L, -M, -MH, and -H -Industrial Ozone Monitors

SHARE

For those who need to integrate their ozone monitor into a separate system such as an air pollution monitoring package, we offer the Model OEM-106 series where the entire ozone monitor is mounted on a single printed circuit board.  The liquid crystal display and power switch may be detached and mounted outside the customer`s custom enclosure.  With the exception of our Personal Ozone Monitor (POM), this is the most compact UV-absorbance-based ozone monitor available on the market.  The Model OEM-106-L has been designated by the EPA as a Federal Equivalent Method (FEM): EQOA–0914–218. Choice of four different optical benches allows for four different ozone concentration ranges as summarized in the following table.

Most popular related searches
  • Measurement Principle: UV Absorption at 254 nm, single beam
  • Measurement Interval: 2 s
  • Data Averaging Options: 10 s, 1 min, 5 min, 1 hr
  • Nominal Flow Rate: ~1 Liter/min; Up to 20 L/min for Model OEM-106-H
  • Data Storage: 32,736 lines (10 s avg. = 3.7 days; 5 min avg = 113 days)
  • Choice of Units: ppb, ppm, pphm, µg m-3, mg m-3 (for Models OEM-106-L and OEM-106-M)
  • ppm, mg m-3 (for Model OEM-106-MH)
  • vol%, wt% air, wt% O2, g/m3, g/Nm3 (for OEM-Model 106-H)
  • Data Outputs: USB, RS232, 0-2.5 V Analog, 4-20 mA, LCD Display
  • Power Requirements: 100-240 V AC, 50/60 Hz
  • 11-28 V DC, nominally 500 mA at 12 V, 6 watt (for Models OEM-106-L, -M, -MH)
  • 11-28 V DC, nominally 325 mA at 12 V, 3.9 watt (for Model OEM-106-H)
  • Size: 2.5 x 7 x 9 inches (6.4 x 17.8 x 22.9 cm)
  • Weight: 2.8 lb (1.3 kg) (for Model OEM-106-L)
    • 2.5 lb (1.1 kg) (for Models OEM-106-M and -MH)
    • 2.9 lb (1.3 kg) (for Model OEM-106-H)
  • Measurement based on UV absorption
  • Low maintenance
  • Low power consumption; can be battery operated
  • Internal data logger with real-time clock
  • 2-s measurement interval
  • On-board microprocessor with interactive menus includes data averaging options of 10 s, 1 m, 5 m and 1 hr
  • USB and RS-232 output of time/date, O3 concentration, internal temperature and pressure
  • Analog output (0-2.5 V and 4-20 mA) of ozone concentration in user-selected units
  • Two 2-level relays for control purposes (e.g., control of ozone source or turn warning light on and off)
  • Long-life pump (15,000 hr) for Models OEM-106-L, OEM-106-M and OEM-106-MH