Teledyne API
  1. Companies
  2. Teledyne API
  3. Products
  4. TAPI - Model 430 - Compact Ozone ...

TAPIModel 430 -Compact Ozone Monitor

SHARE

Teledyne API, a recognized leader in Ozone measurements, is pleased to offer the Model 430 O3 Monitor. The 430 is a compact instrument that uses the proven UV-absorption principle. The small footprint and robust design make it ideal for a variety of applications where portability and accuracy are both necessary.

Most popular related searches
  • Proven UV Absorption bench used in our popular line of Industrial Ozone products.
  • Highest quality of construction.
  • Rugged and lightweight ABS polycarbonate chassis with handle.
  • Well suited for portable, low power monitoring applications.
  • Wide operating range.
  • Ranges : 0 - 100 ppb (min), 0-20,000 ppb (max) (user -selectable)
  • Measurement Units : ppb, ppm, (user- selectable)
  • Zero Noise : < 1 ppb (RMS)
  • Span Noise : < 0.5% of reading (RMS) above 100 ppb
  • Lower Detectable Limit : < 2 ppb
  • Zero Drift : < 1 ppb/24 hours
  • Span Drift : < 1% of full scale/24 hours
  • Lag Time : <20 seconds
  • Rise/Fall Time : < 15 seconds to 95%
  • Linearity : 1% of full scale
  • Precision : < 0.5 ppb or 0.5% of reading above 100 ppb
  • Sample Flow Rate : 0.8 L/min +/-10%
  • Typical Power Consumption : 12VDC, 9W
  • Analog Output Ranges : 0 - 5V
  • Analog Output Resolution : 16-Bit Resolution
  • Internal Data Storage Capacity : 10,000 lines (1 hr avg = 1.1 yr)
  • Included I/O :
    • 1 x RS-232
    • 1 x 0-5V analog output
    • 5 x 0-5V analog input
    • 1 x SD card slot
  • Operating Temperature Range : 5 - 40ºC (with US EPA Approval)
  • Dimensions (HxWxD) : 4.2" x 7.1 ” x 10.2” (107 x 180 x 597 mm)
  • Weight : 5.2 lbs (2.4 kg)
  • Humidity Range : 0-95% RH non-condensing