ENVEA - Model LabAnalyzer 254 - Fast Mercury Determination in Analytical Applications
From Ambient - Air quality monitoring & Laboratory Mercury Instruments
The LabAnalyzer is the perfect solution for quantitative determination of mercury in aqueous samples and sample digests in the laboratory. In contrast to a typical multi-element AAS the LabAnalyzer 254 is specially designed for elemental mercury. This allows top performance in analytical applications. The use of a specially developed highly stable mercury lamp in connection with thermostat-controlled UV sensors results in a detection limit of a few ppt of mercury.
Features & Benefits
- the optical system does not require any readjustment
- Short analysis time
- Low reagent consumption
- Automatic zero adjustment
- Memory effects are minimized and a high sample throughput is possible thanks to specially selected materials for sample gas conducting components and heating of the optical bench.
- For analyses according to DIN 38406-12 / EPA 7470A / EPA 7471A)
- Measuring range:0.01 ppb … 10 ppb (10 ng/l to 10 µg/l)
- Hg trap: Mercury vapors cannot escape into the lab
- Upgradable to VM-3000 Mercury Vapor Monitor
Main Applications
- Water samples: drinking water, waste water, ground water, surface water, sea water
- Soil and sludge samples
- Geological sample material
- Waste samples: glass, construction rumble, contaminated liquids, wood
- Incineration plant monitoring: smoke gas scrubber water, smoke gas analysis (e.g. to VDI 3868-2 VE)
- Monitoring of foodstuffs
- Clinical samples: urine, saliva
- Chemical industry: environmental protection and quality control
- Petrochemical industry
- Scientific research
Technical Specifications
- Measuring principle: cold vapor atomic absorption (CVAAS) at 253.7 nm; Peak method
- UV-Source: Electrodeless low pressure mercury lamp (EDL)
- Stabilization method: Reference beam method
- Optical cell: Fused silica (Suprasil), l: approx. 230mm, heated approx. 45°
- Sample volume: 2 – 10 ml
- Measuring ranges: liquid sample: 0,01 to 10 μg/l at 10 ml sample volume; solid sample: 0,1 ng/g to 0,1 μg/g at 1 g sample weight
- Sensitivity: approx. 5 ng/l resp. 0.05 ng absolutely
- Reducing agent: Tin-II-chloride or sodiumborohydride
Technical Data
- Signal outputs: analogue: 4-20 mA; serial: RS 232 / USB
- Power supply: 230 VAC / 50 Hz (110 VAC / 60 Hz option)
- Power consumption: 100 VA
- Dimensions (W x H x D): Photometer: 45 x 15 x 35 cm; Reactor: 24 x 48 x 27 cm
- Required floor space: approx. 70 x 50 cm (W x D)
- Weight: approx. 10 kg
Approvals & Certifications
- Our products are manufactured according to the ISO 9001 quality regulations.
- CE approved
Customer reviews