- Outstanding precision and speed in hydrogen concentration measurement: The gas analysers are characterized by an industry-leading response time of < 30 ms and a measuring range from a few ppm to 100 vol.%, specially developed for the new requirements of hydrogen infrastructure in the energy and process industries.
- Unrivalled robustness for demanding environments: Optimised for use in humid environments, the devices withstand condensate and water without damage and provides precise measurements up to a pressure of 200 bar – ideal for electrolysers, fuel cells and other hydrogen applications.
- Maximum safety, minimum maintenance: The TCD3000 SiA and TCD3000 Si significantly increase occupational safety through fast and reliable explosion level monitoring. The devices represent a long-term investment with an expected service life of up to 10 years.
- Adaptability meets economy: The screw-in and extremely compact gas analysers save space and costs. Thanks to them, a wide variety of gas mixtures can be measured without additional sample preparation – a highly efficient solution that outperforms the competition in terms of performance and price.
- SiA Dimensions with connectors; Weight - H: 96 mm; Ø: 45 mm; 740 g
- Si Dimensions with connectors; Weight - H: 91 mm; Ø: 45 mm; 440 g
- Gas connection - G 1/2″ or NPT 1/2″
- Power supply - 24 ±25% VDC, < 5 W
- Analog output - 4-20 mA potential-bound, RL ≤ 750 Ω
- Serial interface - RS485, baud rate 38400 / data bits 8
- Ambient temperature range - -40 to +90 °C
- Sample gas temperature - -40 to +90 °C / +125 °C
- Reaction time - ≤ 30 ms
- T90 time - < 1 s
- Warm-up time - < 1 min
- Sample gas pressure - 0,9 to 200 bar (absolute)
- Sample gas flow - 0 to 10 m/s (higher flow rates on request)
- Noise - < 10 ppm
- Measured value drift - < 100 ppm / week
- Repeatability - < 100 ppm
- Temperature influence - < 50 ppm / 10 K
- Influence of sample gas pressure at zero point - < 25 ppm / 10 hPa; ab 1 MPa < 2 ppm / 10 hPa
- Influence of sample gas pressure with deflection gas - < 100 ppm / 10 hPa; ab 1 MPa < 10 ppm / 10 hPa
- Influence of sample gas flow rate at zero point - < 25 ppm / 10 l/h
- Influence of sample gas flow rate with deflection gas - < 100 ppm / 10 l/h
- Multiple calibrations possible for one sensor : Several linearisation curves, different measuring ranges and other gas pairs can be stored on one device (manual switching)
- Pressure compensation : Thermal conductivity is pressure-dependent, therefore a constant pressure must be present during the measurement. If the pressure is not constant, the error can be eliminated by an additional pressure measurement. We can provide a compensation curve on request
- Customised settings possible that go beyond the technical specifications : Customised adjustment for e.g. higher gas temperatures, radioactive environments, particularly low flow rates
- Higher pressure ranges : Settings for high pressure ranges >200 bar also available on request
- Express delivery possible : Depending on capacity and agreement, short-term delivery also possible
- Other connections e.g. NPT ½“ : Standard G 1/2‘ – other connections such as NPT 1/2’ on request
- Can also be ordered as Si variant without ATEX : For use outside the ATEX protection zone