Sonimix - Model 6000 C2 -Multi-Points Multigas Calibrator
FromLNI Swissgas
The multigas Sonimix 6000 C2 calibrator is designed to calibrate manually or by remote control up to 6 monitors installed in an air pollution monitoring station, for parameters such as SO2, HC, BTX, CO, NOx and O3. The Sonimix 6000 C2 includes Mass Flow Controllers, GPT, ozone and self re-generable zero air generator and is equipped with – an inlet for high concentration mixtures of diluted gases such as NO, NO2, SO2, CO, HC, BTX or mixtures of several compatible pollutants in the same cylinder.
Most popular related searches
zero air generator
air contamination monitoring
mass flow control
air pollution monitoring
air pollution monitoring system
gas calibration system
zero air
multi-gas calibration
pollution monitoring system
mass flow
- a zero air generator composed of an internal compressor with self re-generable filters, removes following components: moisture, O3, NOx, SO2, HC, BTX and optionally CO and CH4.
- a binary gas dilutor of two inlets (diluted gas and air), including 2 Mass Flow Controllers to generate the zero and other dilutions ratios going from 1/100 to 1/1000 (other values upon request).
- The input pressures are monitored by two pressure switches. The ozone generator is regulated by UV-light feedback and is compensated from the atmospheric pressure and ambient temperature variations. It includes alarm on the UV power drift as well as the circuits for the Gas Phase Titration (GPT).
- Accuracy – (0,6% rel.)
- No maintenance nor pressurised air cylinder with its included self re-generable zero air generator
- Sonic nozzle technology
- Remote control
- Ergonomic with LCD display
- Air pollution monitoring
Model: Sonimix 6000C2
General data:
- Dilution range: From 1/100 to 1/1000 (50 to 500 ppb with a 50 ppm cylinder)
- Dilution principle: MFC
- Dilution gas: Mixtures of SO2 in Air – NO in N2 – CO in Air – HC or BTX in Air
- Mixtures of several compatible pollutants in the same cylinder
- Diluted gas flow: 0 to 35 Nml/min air
- Dilution gas line flow accuracy: +/- 1.5% relative
- Repeatability on flow: < 0.5% relative
- Stabilization time: < 5 min
- Internal 0 air quality:
- O2 & O3 < 1 ppb – NO & NO2< 2 ppb – HCnm< 50 ppb
- HCm< 50 ppb – CO< 50 ppb – BTEX< 1 ppb
- Remaining humidity level: H2O< 100 ppm (-40°C DP)
- Dimensions:
- 19″ 3HE/84 TE , 500mm deep
- portable casing in option
- Net weight: 20 Kg
Gas inlet requirments:
- Carrier gas: High grade zero air from internal zero air generator
- Flow: 0 to 3500 Nml/min air
- Pressure: 3 +/- 0.3 bar (40 +/- 4 psig)
- Fittings: 1/4’’ and 1/8’’ Swagelok
Gas outlet requirments:
- Flow:: About 3000 Nml/min, other values upon request
- Fittings: 1/4’’ and 1/8’’ Swagelok
O3 section
- O3 concentration: 25 – 500ppb, adjustable by software
- Repeatability on flow: Better than +/- 2,5ppb (at 250ppb O3)
MFC section:
- Linearity and accuracy of the MFC: 0.5 % (rel.) + 0,1 FS
- Repeatability of the MFC: 0.2 % (rel.) + 0,1 FS
Communication:
- RS232: Standard
Electrical data:
- Power supply voltage: 230V/50-60Hz
- 115V/60Hz
- 100V/50-60Hz
- Installed power (Max): 200W