Teledyne API

- Model T200U NOy - Trace Level Chemiluminescence NO-NOy Analyzer

From

0

The Model T200U NOy analyzer has been developed and refined specifically to address the challenges of NOy monitoring as required, for example, in the US NCore network. It uses the proven chemiluminescence principle and is coupled with a remote NOy converter via umbilical to allow measurements with a lower detectable limit of 50 ppt.

The Model T200U NOy analyzer combines a gold plated reaction cell with an enhanced performance pump, special low noise photomultiplier tube, and unique signal conditioning to provide exceptional sensitivity. In addition, a pre-reactor separates the NO-O3 reaction from background chemiluminescence to allow accurate auto-zero of the analyzer. The remote NOy converter is provided with a stainless steel, NEMA 4X (IP65) enclosure and custom sample umbilical to allow a mounting height of 10 meters. In addition, the converter features a unique purge circuit designed to remove residual calibration gas.

All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability. All instrument set up, control and access to stored data and diagnostic information is available through the front panel, or via RS232, Ethernet, or USB com ports, either locally or by remote connection.

The Model T200U NOy comes with the Numaview™ Software. NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView™ Software.

  • Ranges: 0-5 ppb to 0-2,000 ppb, user selectable
  • Independent ranges and auto ranging
  • Large, vivid, and durable color graphics display with touch screen interface
  • Ethernet, RS-232, and (optional) USB com ports
  • Front panel USB connections for peripheral devices and firmware upgrades
  • 8 analog inputs (optional)
  • Adaptive signal filtering optimizes response time
  • Temperature & pressure compensation
  • 50 ppt lower detectable limit
  • Catalytic ozone scrubber
  • Comprehensive internal data logging with programmable averaging periods
  • Ability to log virtually any operating parameter
  • Two-year warranty
  • External converter not shown above

  • Ranges :
    • Min: 0-5 ppb Full scale
    • Max: 0-2,000 ppb Full scale (selectable, independent NO, NOy ranges)
  • Measurement Units : ppb, μg/m3 (selectable)
  • Zero Noise : < 25 ppt (RMS)
  • Span Noise : < 0.5% of reading (RMS) above 5 ppb
  • Lower Detectable Limit : <50 ppt
  • Zero Drift : < 0.1 ppb/24 hours
  • Span Drift : < 0.5% of reading/24 hours
  • Lag Time : 20 seconds
  • Rise/Fall Time : < 50 seconds to 95%
  • Linearity : 1% of full scale
  • Precision : 0.5% of reading above 5 ppb
  • Sample Flow Rate : 1800 cm3/min ±200 cm3/min (NO & NOy) (Includes 800 cc bypass)
  • Power Requirements : 100V-120V, 220V-240V, 50/60 Hz
  • Analog Output Ranges : 10V, 5V, 1V, 0.1V (selectable)
  • Recorder Offset : ±10%
  • Included I/O :
    • 1 x Ethernet: 10/100Base-T
    • 2 x RS232 (300-115,200 baud)
    • 2 x USB device ports
    • 8 x opto-isolated digital outputs
    • 6 x opto-isolated digital inputs
    • 4 x analog outputs
  • Optional I/O :
    • 1 x USB com port
    • 1 x RS485
    • 8 x analog inputs (0-10V, 12-bit)
    • 4 x digital alarm outputs
    • Multidrop RS232
    • 3 x 4 - 20mA current outputs
  • Operating Temperature Range : 5 - 40ºC
  • Dimensions (HxWxD) : Analyzer: 7” x 17” x 23.5” (178 x 432 x 597 mm)
  • Flow Module: 8.75” x 17” x 23.5” (222 x 432 x 597 mm)
  • Converter: 13.5” x 10.5” x 6.5” (343 x 267 x 165 mm)
  • Weight :
    • Analyzer: 40 lbs (18 kg)
    • Flow Module: 28 lbs (12.7 kg)
    • Converter: 15 lbs (6.8 kg)
    • External pump: 21 lbs (9.5 kg)

Customer reviews

No reviews were found for TAPI - Model T200U NOy - Trace Level Chemiluminescence NO-NOy Analyzer​. Be the first to review!