Solinst Canada Ltd.

SolinstModel 3500 AquaVent 5 -Vented Water Level Datalogger

SHARE
The AquaVent 5 uses the same vented pressure transducer and Vented Cable as the LevelVent 5; however, the batteries are user-replaceable and located in the AquaVent 5 Wellhead at surface. The AquaVent 5 Wellhead provides options to communicate with Solinst Levelogger Software and accessories as well as third party equipment using SDI-12 or MODBUS protocols. The Wellhead fits conveniently onto a 50 mm (2") well casing. Permanent desiccants and hydrophobic filters protect the AquaVent 5 from moisture.
Most popular related searches

The AquaVent 5 Vented Water Level Datalogger records accurate water level and temperature measurements in shallow groundwater and surface water applications. It combines pressure and temperature sensors, hydrophobic filters and datalogger within a 22 mm x 173 mm (7/8" x 6.8") stainless steel housing with corrosion-resistant coating.

The AquaVent 5 uses a gauged pressure transducer; it is open to the atmosphere via a vented cable to surface. Atmospheric pressure is applied to the transducer diaphragm, providing a cancellation effect for barometric pressure. This results in actual water level recordings.

The vented cable and Solinst AquaVent 5 logger are protected from moisture by built-in desiccants and hydrophobic filters.

The vented pressure transducer is made of Hastelloy®, making it extremely durable and accurate in a wide range of temperature and monitoring conditions. The sensor provides an accuracy of 0.05% FS, and can withstand 2 times over-pressure without permanent damage.

The robust memory can hold up to 150,000 data sets when programmed in Solinst Levelogger® Software.

  • Ideal for shallow applications: up to 65 ft (20 m) submergence
  • Aquifer characterization: pumping tests, slug tests, etc.
  • Stream gauging, lake and reservoir management
  • Watershed, drainage basin and recharge monitoring
  • Stormwater and runoff monitoring
  • Long-term water level monitoring in wells and surface water
  • Gauged pressure sensor for highly-accurate water level measurements: 0.05% FS
  • Multiple built-in hydrophobic filters and desiccants - no need to replace, reducesAquaVent maintenance
  • Easy-to-access, user-replaceable batteries in Wellhead
  • Options for MODBUS (RS-232/RS-485) and SDI-12
  • Separate cables for each communication protocol
  • Corrosion and abrasion-resistant coating baked-on using polymerization technology
  • Automatic barometric compensation reduces time required for post data processing
  • Integrate into a third-party data collection system for remote real-time data
  • Continuous, reliable water level data for long-term monitoring projects
  • Actual water level readings for instant aquifer test results
  • Level Sensor : Piezoresistive Silicon with Hastelloy Sensor
  • Accuracy : ± 0.05% FS Typical
  • Stability of Readings : Superior, low noise
  • Units of Measure : m, cm, ft., psi, kPa, bar, ºC. ºF
  • Normalization : Automatic Temperature Compensation
  • Temp. Comp. Range : 0º to 50ºC
  • Temperature Sensor : Platinum Resistance Temperature Detector (RTD)
  • Operating Temperature : -20ºC to 80ºC
  • Temp. Sensor Accuracy : ± 0.05ºC
  • Temp. Sensor Resolution : 0.003°C
  • Battery Life : 8 Years - based on 1 reading/minute
  • Clock Accuracy : ± 1 minute/year (-20ºC to 80ºC)
  • Maximum # Readings : 150,000 sets of readings
  • Memory : Slate and Continuous
  • Communication : 57600 bps Solinst USB, 1200 baud SDI-12, Modbus RS-485/RS-232 (various speeds)
  • Size : 7/8" x 6.8" (22 mm x 173 mm)
  • Weight : 6.4 oz. (182 grams)
  • Corrosion Resistance: Baked-on coating using polymerization
    technology (inside and out)
  • Wetted Materials : Delrin®, Viton®, 316L stainless steel, Hastelloy, PFAS-free PTFE coating
  • Sampling Modes : Linear, Event & User-Selectable with Repeat Mode, Future Start, Future Stop, Real-Time View
  • Measurement Rates : 1/8 sec to 99 hrs
  • Barometric Compensation : Automatic